Evaluation of a subcutaneous continuous glucose monitoring system in critically ill neonatal foals

View Full Paper
FPFlavie PayetteLLL. LeducDSDarko Stefanovski

Key Points

  • To evaluate the accuracy and correlation of continuous glucose monitoring systems relative to standard measurement methods in neonatal foals.
  • Conducted a prospective method comparison study with 15 critically ill neonatal foals.
  • Measured glucose concentrations using CGMS, point-of-care glucometry, and laboratory analysis at defined intervals.
  • Assessed blood pressure to evaluate hypotension effects on CGMS readings.
  • Average bias between CGMS and laboratory analysis was 48 mg/dL, indicating significant discrepancies.
  • Spearman’s correlation coefficients revealed moderate agreements between measurement methods.
  • CGMS showed only 15.5% accuracy within 15% of laboratory values compared to 88.3% for point-of-care glucometry.

Abstract

Abstract Background Dysglycemia is common in critically ill neonatal foals. Continuous glucose monitoring systems (CGMS) are potentially useful in these cases, but factors such as poor peripheral perfusion could interfere with results. Hypothesis/Objectives Evaluate the correlation, agreement, and accuracy of CGMS compared to point-of-care glucometry (POCG) and laboratory analysis (LAB) in critically ill neonatal foals and assess the impact of hypotension on CGMS measurements. Animals Fifteen critically ill, client-owned neonatal foals. Methods In a prospective method comparison study utilizing clinical cases, glucose concentration was measured serially using CGMS and POCG (every 6 h), and LAB (every 24 h) for pairwise comparison. Blood pressure was measured every 12 h. Results Average bias (95% limits of agreement) between CGMS and LAB, POCG and LAB, and CGMS and POCG were 48 mg/dL (−27 to 111), 10 mg/dL (−23 to 45), and 38 mg/dL (−21 to 98), respectively. Spearman’s correlation was significant between CGMS and LAB (r = 0.65), POCG and LAB (r = 0.77), and CGMS and POCG (r = 0.75). The CGMS accuracy was low with only 15.5% of CGMS concentrations within 15% of LAB concentrations, compared with 88.3% for POCG. Hypotension did not affect CGMS measurements. Conclusions and clinical importance The CGMS provided glucose measurements above both LAB and POCG concentrations. Given its low correlation and accuracy, CGMS cannot replace LAB or POCG as the sole glucose measurement method in critically ill foals. However, it is a useful adjunct for tracking trends and providing alerts.