Compatibility and Bias Between Ultrasound-guided Attenuation Parameter and Controlled Attenuation Parameter for Ultrasound-based Hepatic Steatosis Assessment

View Full Paper
DMDaisuke MiuraRHRino HiwatashiTSTomoko Sakata

Key Points

  • This study aims to develop a regression model estimating controlled attenuation parameter (CAP) from ultrasound-guided attenuation parameter (UGAP) values and assess their interchangeability.
  • Performed linear regression analyses to derive predictive equations for CAP using UGAP-AC.
  • Conducted Bland-Altman analysis to evaluate agreement and biases between UGAP and CAP.
  • Utilized repeated-measures analysis of variance to assess the impact of body mass index (BMI) on measurement differences.
  • UGAP-AC and CAP demonstrated a strong positive correlation (r = 0.947, P < 0.0001).
  • Developed regression equation: CAP (dB/m) = 371.5 × UGAP-AC + 12.3 (R² = 0.8946).
  • Fixed bias of 28.6 dB/m was identified; BMI influenced the proportional bias but was less relevant in the diagnostic CAP range (228 to 300 dB/m).

Abstract

This retrospective observational study aimed to develop a regression model to estimate the controlled attenuation parameter (CAP) from the ultrasound-guided attenuation parameter (UGAP) attenuation coefficient (AC), and to evaluate the agreement, biases, and potential interchangeability between the 2 measurements in a routine clinical setting. A total of 85 patients who underwent both UGAP and CAP measurements on the same day were included. UGAP-AC values were converted from dB/cm/MHz to dB/m to match CAP units. Pearson’s correlation and linear regression analyses were performed to derive a predictive equation for CAP based on UGAP-AC. Agreement and biases between the 2 attenuation values were assessed using Bland-Altman analysis. In addition, repeated-measures analysis of variance was conducted to assess the influence of body mass index (BMI) on measurement differences. UGAP-AC and CAP showed a strong positive correlation ( r =0.947, P <0.0001). The regression equation was CAP (dB/m)=371.5× UGAP-AC + 12.3 (R²=0.8946). Bland-Altman analysis revealed a fixed bias of 28.6 dB/m and a BMI-dependent proportional bias. However, this proportional bias disappeared observed when analysis was limited to the clinically relevant CAP range (228 to 300 dB/m), suggesting good comparability between UGAP and CAP within the diagnostic range. UGAP-AC can reliably estimate CAP using a simple linear regression model. While some systematic biases exist, their clinical impact is limited. These findings support the potential interchangeability of UGAP and CAP for noninvasive hepatic steatosis assessment in high-throughput clinical practice.