February 5, 2026Aesthetic Surgery Journal0 citations

Silicone Axillary Lymphadenopathy Associated With Rupture of Fourth-Generation Cohesive Silicone Breast Implants Following Breast Augmentation

View Full Paper
SBStephen D. Bresnick

Key Points

  • To evaluate the occurrence of silicone axillary lymphadenopathy (SAL) in relation to ruptured fourth-generation cohesive silicone breast implants.
  • Conducted a retrospective review involving 260 patients with confirmed ruptured silicone implants.
  • Utilized non-contrast MRI and ultrasound to identify implant rupture.
  • Correlated imaging findings with operative results and implant characteristics, including age and type.
  • Analyzed survival of both intracapsular and extracapsular rupture types against axillary findings.
  • Identified 299 ruptured implant-sides among the patients, with surgical findings confirming imaging results.
  • Intracapsular ruptures accounted for 81.6% while extracapsular ruptures comprised 18.4%.
  • SAL was more frequently associated with extracapsular ruptures than intracapsular.
  • A significant link was determined between increasing implant age at explantation and the incidence of SAL.

Abstract

Abstract Background Silicone implant rupture is a common complication of breast augmentation, with silicone axillary lymphadenopathy (SAL) a recognized sequela. Objectives To evaluate SAL associated with surgically confirmed rupture of fourth-generation cohesive silicone breast implants and correlate imaging with operative findings. Methods A retrospective review included 260 consecutive patients undergoing surgery for removal of one or more ruptured cohesive silicone implants identified by non-contrast breast/axillary MRI and ultrasound. The implant-side was the unit of analysis, with bilateral ruptures counted separately. Implant characteristics, including manufacturer, volume, rupture type, and implant age at explantation, were correlated with ipsilateral axillary findings. All patients had primary augmentation with Allergan plc (Irvine, CA) or Mentor Worldwide LLC (Irvine, CA) fourth-generation cohesive implants and available implant records. Operative findings confirmed implant type, size, rupture presence, and rupture characteristics. Results Among 260 patients, 299 ruptured implant-sides were identified, with imaging findings corroborated surgically. Intracapsular rupture accounted for 81.6% of ruptured implant-sides with 18.4% extracapsular ruptures. SAL occurred with both intracapsular and extracapsular rupture types but was more commonly seen with extracapsular rupture. Increasing implant age at explantation showed a strong independent association with SAL, while extracapsular rupture demonstrated a modest independent association. No significant differences in SAL were observed by implant manufacturers. Conclusions SAL was frequently observed in patients undergoing surgical management of ruptured smooth, fourth-generation cohesive silicone implants. Increasing implant age at explantation was independently associated with SAL, while extracapsular rupture showed a modest independent association. These findings strongly influence patient counseling and surgical evaluation of aging fourth-generation silicone implants. Further study of newer fifth-generation highly cohesive implants is warranted.