February 6, 2026Bioengineering0 citationsOpen Access

Harmonic Ratio Analysis in Magnetic Particle Imaging Enables Differentiation of Malignant and Benign Human Breast Tissues: A Feasibility Study

View Full Paper
HYHongyu YangHZHaoran ZhangYZYiyin Zhang

Key Points

  • This research aims to evaluate the capability of harmonic ratio analysis in differentiating malignant and benign breast tissues using magnetic particle imaging.
  • Utilized magnetic particle imaging to measure harmonic response in breast tissue samples.
  • Applied Synomag-50 nanoparticles to enhance magnetic response in tissue samples.
  • Implemented a custom-built MPI spectrometer with rapid measurement protocol.
  • Calculated the harmonic ratio of the third to fifth harmonics (H3/H5).
  • Malignant tissue samples exhibited harmonic ratios exceeding 2.2.
  • Benign samples' harmonic ratios remained below 2.0.
  • A stepwise increase in harmonic ratios was noted from normal to tumor-adjacent, then to malignant tissues.
  • Histological analysis confirmed effective infiltration of magnetic nanoparticles.

Abstract

Accurate intraoperative differentiation between malignant and benign breast tissues, particularly the assessment of lymph node status and tumor margins, is critical for surgical decision-making and prognosis. Traditional histopathological methods, such as frozen section analysis, are time-consuming and labor-intensive. Magnetic Particle Imaging (MPI) is a novel, radiation-free modality that senses the microenvironmental properties of tissues through the dynamic response of magnetic tracers. In this study, we propose a diagnostic method utilizing the higher-order harmonic response of magnetic nanoparticles. Various ex vivo breast tissue samples were immersed in Synomag-50 nanoparticles. Using a custom-built MPI spectrometer (5 kHz excitation, 9 mT amplitude) operating in spectroscopic mode, we implemented a rapid acquisition protocol in which each sample was measured 10 times, with 0.1 s per cycle. We analyzed the magnetic response spectrum and calculated the ratio of the third to the fifth harmonic (H3/H5). Histological analysis confirmed the effective infiltration of MNPs into the interstitial spaces. The repeated measurement data demonstrated high stability. A distinct stepwise increase in harmonic ratios was observed from normal tissue to tumor-adjacent tissue and finally to malignant tumors. Specifically, malignant samples showed ratios that generally exceeded 2.2, whereas benign samples remained below 2.0. These preliminary findings suggest that the harmonic ratio could serve as a sensitive biomarker reflecting the microenvironmental constraints associated with malignancy. This study validates the feasibility of utilizing MPI signal harmonics as a quantitative metric with rapid signal acquisition capabilities for differentiating benign and malignant lymph nodes.