Study of the over-response of radiochromic films in ultra-high dose rate and dose per pulse electron beams

View Full Paper
FHFelix HorstASAndreas SchüllerRKRalf‐Peter Kapsch

Key Points

  • The research aims to determine if the over-response of EBT3 films observed with protons occurs in UHDR electron radiation.
  • Irradiated around 600 EBT3 samples using UHDR electron beams at two different facilities.
  • Compared EBT3 films with OC-1 films, which are dose rate independent.
  • Used PTW flashDiamond detector and alanine dosimetry as dosimetric references.
  • EBT3 films showed an over-response of up to 50% at higher doses per pulse in UHDR conditions.
  • No significant deviations were found for doses up to 3 Gy per pulse.
  • OC-1 films did not demonstrate any significant effects under the same conditions.

Abstract

Abstract Objective: Many pre-clinical experiments on FLASH radiotherapy, in which biological samples are irradiated at conventional and ultra-high dose rate (UHDR), make use of radiochromic films as an absolute dose reference because their response is typically assumed to be dose rate independent. While several previous experiments confirmed this assumption, a recent study with protons found an over-response for Gafchromic EBT3 films at UHDR doi:10.1002/mp.15526. If EBT3 is used as a reference in FLASH experiments, the over-response could result in an underdosing that falsely appears as sparing of normal tissue.Approach: To investigate whether the reported dose rate dependence of EBT3 can be reproduced with UHDR electron radiation as well, about 600 EBT3 samples were irradiated at the metrological linar accelerator facility at Physikalisch Technische Bundesanstalt, Braunschweig, and at the ELBE research accelerator at Helmholtz-Zentrum Dresden-Rossendorf, with different beam parameters. As dosimetric reference served a PTW flashDiamond detector which has been verified to show no deviation in the studied UHDR range as well as PTB's alanine dosimetry system. In addition, several OC-1 films, which have been shown to be dose rate independent in previous studies, were irradiated together with EBT3 films.Main results: EBT3 shows a significant over-response at ultra-high dose per pulse and intra-pulse dose rates. Up to 3 Gy per pulse no significant deviations were observed.However, for larger doses per pulse an over-response was observed that increases with dose per pulse to up to 50%. For OC-1 films no significant effects were found.Significance: We provide evidence for the over-response of EBT3 films in UHDR electron beams. Therefore, EBT3 films should be used with caution in FLASH experiments.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Horst et al. (2026) studied this question.

synapsesocial.com/papers/69994b88873532290d01fb2ehttps://doi.org/10.1088/1361-6560/ae47b6