March 1, 2026Journal of Biomedical Optics0 citationsOpen Access

Cause, effect, and remediation of melanin-associated bias in pulse oximetry

KBKevin BennerNGNeha GoelMRMark S. Rea

Key Points

  • The research aims to investigate melanin's impact on pulse oximetry accuracy and explore how changing light sources can mitigate hypoxemia risk.
  • Analyzed published mean errors and uncertainties in SpO2 readings for Black and White patients.
  • Developed an analytical model to predict hypoxemia risk based on skin melanin levels.
  • Compared errors from current and hypothetical monochromatic light oximeters.
  • Predicted risk of occult hypoxemia for Black patients was 2.1 times higher than for White patients.
  • Using monochromatic light sources would reduce mean errors in readings by 83% for Black patients and 65% for White patients.
  • Findings support previous empirical research on pulse oximeter inaccuracies related to skin melanin.

Abstract

SignificanceBlack patients are at greater risk than White patients for occult hypoxemia due to a melanin-associated bias in SpO2 readings from commercially available pulse oximeters that employ polychromatic light sources.AimWe aim to demonstrate how melanin-associated inaccuracies in commercially available pulse oximeters increase the likelihood of occult hypoxemia for all patients and how monochromatic light sources could minimize melanin-associated occult hypoxemia.ApproachPublished values of mean error (M SpO2−SaO2) and uncertainty (SD SpO2−SaO2) for Black and White patients were used to analytically model the risk of occult hypoxemia. Mean errors and uncertainties were also estimated for hypothetical patients with no skin melanin, which are directly comparable to values for pulse oximeters employing monochromatic light sources.ResultsThe analytically predicted risk of occult hypoxemia for Black patients relative to White patients was 2.1, consistent with published empirical findings. Oximeters unaffected by melanin in the skin would have smaller mean errors and uncertainties than current oximeters, significantly reducing the risk of occult hypoxemia in both Black patients (83% reduction) and White patients (65% reduction).ConclusionsBecause everyone has melanin in their skin, the use of monochromatic light sources in pulse oximeters could significantly reduce the risk of occult hypoxemia for everyone.