March 3, 2026Medical Devices Evidence and Research0 citationsOpen Access

A Benchtop Evaluation of Cervical Collar Design and Strap Tension

View Full Paper
LRLaurence RussellLJLiudi JiangDFDavide Filingeri

Key Points

  • Interface pressure values exceeded 100 mmHg for high strap tensions, indicating potential discomfort risks.
  • A significant correlation of r > 0.4 was found between strap tension and interface pressure across the evaluated collars.
  • This observational analysis evaluated four commercially available cervical collars using an innovative experimental setup.
  • Findings highlight the need for objective pressure measurements in assessing cervical collar designs and fit.

Abstract

Abstract: This article presents a novel experimental setup for objectively evaluating cervical collar design and fit by monitoring strap tension and interface pressures across the device-skin interface. This study evaluated four commercially available cervical collars, achieving repeatable findings (ICC > 0.93) that were sensitive to differences between collars and across strap tensions. For each collar, there was a significant relationship (r > 0.4, p < 0.005) between strap tension and interface pressure values, with pressures exceeding 100 mmHg for high strap tensions. Differences in the distribution of pressures and the relationship between strap tension and interface pressures highlight the need for these objective measures of design and fit for cervical collars to evaluate new and existing designs. Keywords: cervical collars, pressure ulcers, interface pressure

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Russell et al. (2026) studied this question.