Bench Validation of Low-Cost Digital Manometry for Intraoral Pressure Measurements Relevant to Aerodynamic Assessments of Voice

View Full Paper
JCJ. Randall CurtisASApril SilvestreKKKate Kocins

Key Points

  • This study aims to validate pressure measurements of a low-cost digital manometer in aerodynamic voice assessments.
  • Laboratory-based study design
  • Used a motorized air pump and simultaneous pressure testing
  • Compared pressure measurements between PAS and LDM
  • Assessed concurrent validity using statistics including Lin's concordance and Bland-Altman plots
  • Observed substantial agreement between the LDM and PAS
  • Bland-Altman analysis indicated consistent negative bias in LDM measurements
  • Regression analysis showed excellent predictive accuracy (R² = .999)

Abstract

Purpose: Aerodynamic assessments of voice are part of the American Speech-Language-Hearing Association–recommended, comprehensive approach for evaluating voice and voice disorders. However, one of the most widely used commercial multichannel systems for aerodynamic assessment, the Phonatory Aerodynamic System (PAS), is expensive and cumbersome, limiting widespread use. To address this clinical barrier, the aim of this proof-of-concept study was to assess the validity of pressure measurements using a commercially available, low-cost digital manometer (LDM). Method: This laboratory-based study used a motorized air pump and a simultaneous pressure testing study design to compare differences in pressure measurement between the PAS and the LDM. Concurrent validity was assessed using descriptive statistics, Lin's concordance correlation coefficient (ρ c ), Bland–Altman plots, and linear regression models. Concordance was interpreted as “poor” if ρ c < .90, “moderate” if ρ c = .90 to < .95, “substantial” if ρ c = .95 to < .999, and “excellent” if ρ c ≥ .999. Results: Substantial agreement was observed between the LDM and the PAS. Bland–Altman analysis revealed a consistent negative bias, with the LDM underestimating pressure by approximately 0.36 cmH 2 O. Regression modeling demonstrated excellent predictive accuracy ( R 2 = .999), with correction improving alignment between devices. Conclusions: The findings from this study demonstrate the potential for LDMs to enhance the clinical feasibility of obtaining intraoral pressure measurements that contribute to aerodynamic voice assessments, particularly in settings where access to comprehensive multichannel instrumentation is limited. Future research will compare aerodynamic measures of voice using LDM and PAS with human participants during phonatory tasks, expanding upon these proof-of-concept laboratory-based findings.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Curtis et al. (2026) studied this question.

synapsesocial.com/papers/69b4ba1818185d8a39802a28https://doi.org/10.1044/2025_persp-25-00260