Enhanced Sensitivity and Isomer Differentiation of Alkyl Nitrites Using a Pulsed DC SPI-MS

View Full Paper
YNYoko NunomeAFAyano FujiiCSChika Shimabukuro

Key Points

  • This research aims to develop a highly sensitive analytical method for detecting alkyl nitrites and differentiating their isomers.
  • Utilized pulsed dc soft plasma ionization-quadrupole mass spectrometry (SPI-MS) system.
  • Operated using ambient air at controlled pressure to enhance sensitivity.
  • Adjusted duty ratio to manipulate ion-molecule reaction pathways.
  • Achieved a limit of detection (LOD) of 0.16 parts per million by volume (ppmv).
  • Enhanced signal intensity and target ion occupancy by decreasing duty ratio from 50% to 5%.
  • Enabled rapid identification of C4–C6 alkyl nitrite isomers.

Abstract

Despite their significance as forensic targets, alkyl nitrites, classified as illegal drugs, have received little attention in forensic analysis due to their high volatility and chemical instability. Here, we present a high-performance analytical approach using a pulsed dc soft plasma ionization-quadrupole mass spectrometry (pulsed dc SPI-MS) system, uniquely designed to operate using ambient air as the discharge gas. In this system, the modulation of the duty ratio functions as a “structural probe” to identify reactive isomers. Unlike conventional dielectric barrier discharge (DBD) sources that typically operate at atmospheric pressure, our SPI system utilizes a controlled pressure regime of several kPa, where the nitrogen in the ambient air effectively functions as a third-body gas to suppress excessive internal energy. The control of the duty ratio in our pulsed dc SPI source allowed for the successful manipulation of ion–molecule reaction pathways for highly reactive analytes. By optimizing several parameters, including duty ratio and discharge pressure, we achieved a unique ionization regime where the molecular-related ion 2 M − 3 H+ was predominantly detected as the base peak with minimal fragmentation. Notably, by reducing the duty ratio from 50% to 5%, both the target ion occupancy and signal intensity were significantly enhanced, achieving a limit of detection (LOD) as low as 0.16 parts per million by volume (ppmv). This sensitivity is several orders of magnitude higher than previously reported thresholds, enabling rapid identification of C4–C6 alkyl nitrite isomers. This method transforms the duty ratio into a powerful diagnostic tool for identifying reactive intermediates, providing a practical and efficient approach for the onsite identification of illegal alkyl nitrites in forensic and security fields.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Nunome et al. (2026) studied this question.

synapsesocial.com/papers/69cd7b065652765b073a8ae6https://doi.org/10.3390/appliedchem6020020
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Tuning Soft Ionization Strength for Organic Mass Spectrometry2016 · 21 citations
  2. 2Solvent-free sample preparation by headspace solid-phase microextraction applied to the tracing of n-butyl nitrite abuse1996 · 15 citations
  3. 3Gas discharge plasmas and their applications2002 · 1,089 citations
  4. 4Gas phase ion-molecule reactions of alkyl nitrites by ion cyclotron resonance spectroscopy1978 · 22 citations
  5. 5The study of the pyrolysis of alkyl nitrites in a field ionization mass spectrometer1981 · 6 citations