PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
December 1, 2008Circulation Arrhythmia and Electrophysiology633 citations

Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus

View Full Paper
KYKatsuaki YokoyamaElectrophysiologyHNHiroshi NakagawaElectrophysiologyDSDipen ShahElectrophysiology

Structured PICO

Does a novel contact force sensor in an irrigated radiofrequency ablation catheter predict lesion size and the incidence of steam pop and thrombus?

I
Intervention
Open-irrigated radiofrequency ablation catheter with contact force sensor
O
Outcome
Radiofrequency lesion size and incidence of steam pop and thrombussurrogate

A novel contact force sensor in an open-irrigated ablation catheter predicts lesion size and complication risks, potentially optimizing radiofrequency power and application time.

Abstract

This novel ablation catheter, which accurately measures CF, confirmed CF is a major determinant of RF lesion size. Steam pop and thrombus incidence also increases with CF. CFS in an open-irrigated ablation catheter that may optimize the selection of RF power and application time to maximize lesion formation and reduce the risk of steam pop and thrombus.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yokoyama et al. (2008) studied this question.

synapsesocial.com/papers/69d56ec275589c71d767d5f4https://doi.org/10.1161/circep.108.803650
Ask AI
Helpful
Bookmark
Share
View Full Paper