April 10, 2026Healthcare0 citationsOpen Access

Usability of a Patch-Type Ultrasound System for Non-Invasive Hemodynamic Monitoring: A Simulation Study in Anesthesiologists

SNS. NohHKHo KimHCHyeonkyeong Choi

Key Points

  • This research aims to evaluate the usability of a patch-type ultrasound system for hemodynamic monitoring in clinical settings.
  • Conducted a summative evaluation per IEC 62366-1 guidelines with 15 anesthesiologists.
  • Identified potential hazards using the FDA MAUDE database to inform test scenarios.
  • Participants were stratified by clinical experience and performed 9 clinical scenarios with 49 tasks.
  • Measured task success rates, subjective satisfaction using a Likert scale, and the System Usability Scale.
  • Overall task success rate was 98.2%.
  • No significant differences in performance or satisfaction across experience levels (p > 0.05).
  • Mean System Usability Scale score was 78.5, indicating good usability levels.
  • Some user errors were linked to prior device experiences rather than complexity of the interface.

Abstract

Background/Objectives: Non-invasive hemodynamic monitoring technologies are being developed to support clinical decisions while reducing risks from invasive procedures. Usability evaluation is essential to assess safety and effectiveness before commercial release. This study examined the usability of a novel patch-type ultrasound-based system (CW10) designed for continuous monitoring in perioperative settings. Methods: A summative evaluation was conducted following IEC 62366-1 with 15 anesthesiologists. Potential hazards were identified via the FDA MAUDE database (Code: DQK) to inform test scenarios. Participants were stratified by clinical experience (1– 0.05). The mean SUS score was 78.5, corresponding to a “Good” usability level. While some use errors occurred in tasks like probe orientation, root cause analysis suggested these were likely due to negative transfer from prior device experience rather than interface complexity. Conclusions: The results suggest the system demonstrates acceptable usability and consistent operation across experience levels. Integrated automated features and the patch design may contribute to reducing inter-user variability for continuous monitoring. This study provides usability evidence that may inform the development of similar non-invasive technologies.