July 9, 2004Clinical Oral Implants Research372 citations

Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading

View Full Paper
RGRoland GlauserLSLars SennerbyNMNeil Meredith

Key Points

Key points are not available for this paper at this time.

Abstract

Within the limitations of this study, it is concluded that failing implants show a continuous decrease of stability until failure. Low RFA levels after 1 and 2 months seem to indicate an increased risk for future failure. This information may be used to avoid implant failure in the future by unloading implants with decreasing degree of stability with time as diagnosed with the RFA technique.