April 1, 1983Clinical Orthopaedics and Related Research237 citations

Biomechanical Properties of Bone Allografts

View Full Paper
RPRichard R. PelkerGFGary E. FriedlaenderTMThomas C. Markham

Key Points

Key points are not available for this paper at this time.

Abstract

The biomechanical properties of allograft bone can be altered by the methods chosen for its preservation and storage. These effects are minimal with deep-freezing or low-level radiation. Freeze-drying, however, markedly diminishes the torsional and bending strength of bone allografts but does not deleteriously affect the compressive or tensile strength. Irradiation of bone with more than 3.0 megarad or irradiation combined with freeze-drying appears to cause a significant reduction in breaking strength. These factors should be considered when choosing freeze-dried or irradiated allogeneic bone that will be subjected to significant loads following implantation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Pelker et al. (1983) studied this question.

synapsesocial.com/papers/69da2a9b0d540cafc5838c32https://doi.org/10.1097/00003086-198304000-00009