November 8, 2021Journal of Medical Internet Research158 citationsOpen Access

Accuracy Assessment of Oura Ring Nocturnal Heart Rate and Heart Rate Variability in Comparison With Electrocardiography in Time and Frequency Domains: Comprehensive Analysis

View Full Paper
RCRui CaoIAIman AzimiFSFatemeh Sarhaddi

Key Points

  • The aim was to evaluate the accuracy of Oura Ring measurements for nocturnal heart rate and heart rate variability against electrocardiography.
  • Tested Oura Ring against electrocardiography measurements in nocturnal settings.

Structured PICO

Does the Oura Ring accurately measure nocturnal heart rate and heart rate variability compared to electrocardiography?

I
Intervention
Oura Ring
C
Comparator
Electrocardiography (ECG)
O
Outcome
Accuracy of nocturnal heart rate (HR) and heart rate variability (HRV) metrics (RMSSD, AVNN, pNN50, HF, SDNN, LF, LF:HF ratio)surrogate

The Oura Ring provides accurate nocturnal HR and RMSSD measurements but is unreliable for low-frequency HRV metrics compared to standard ECG.

Abstract

The Oura Ring could accurately measure nocturnal HR and RMSSD in both the 5-minute and average-per-night tests. It provided acceptable nocturnal AVNN, pNN50, HF, and SDNN accuracy in the average-per-night test but not in the 5-minute test. In contrast, the LF and LF:HF ratio of the ring had high error rates in both tests.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Cao et al. (2021) studied this question.

synapsesocial.com/papers/69da8b9a85037e71b26840dchttps://doi.org/10.2196/27487