November 12, 2013Journal of Neural Engineering626 citationsOpen Access

Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates

View Full Paper
JBJames C. BarreseNRNaveen G. RaoKPKaivon Paroo

Key Points

Key points are not available for this paper at this time.

Abstract

This is the first long-term failure mode analysis of an emerging BCI technology in a large series of non-human primates. The classification system introduced here may be used to standardize how neuroprosthetic failure modes are evaluated. The results demonstrate the potential for these arrays to record for many years, but achieving reliable sensors will require replacing connectors with implantable wireless systems, controlling the meningeal reaction, and improving insulation materials. These results will focus future research in order to create clinical neuroprosthetic sensors, as well as valuable research tools, that are able to safely provide reliable neural signals for over a decade.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Barrese et al. (2013) studied this question.