April 12, 2026Journal of Applied Clinical Medical Physics0 citationsOpen Access

Assessment of LD‐V1 radiochromic film low‐dose performance for mega‐voltage radiotherapy quality assurance

View Full Paper
RERuiming C. EdmondsonDGD. GarciaVMVictor Malkov

Key Points

  • This research evaluates the performance of LD-V1 radiochromic film for low-dose measurements in mega-voltage radiotherapy quality assurance.
  • Comparison of LD-V1 film with EBT4 for low-dose measurements.
  • Assessment of low-dose sensitivity under various conditions.
  • Evaluation of energy and scanning-orientation effects.
  • LD-V1 exhibits superior low-dose sensitivity compared to EBT4.
  • Significant variations in performance based on energy levels and scanning orientation are observed.
  • Consistent calibration and handling are critical for optimal accuracy.

Abstract

LD-V1 demonstrates suitability for low-dose MV measurements and exhibits improved low-dose sensitivity relative to EBT4, indicating its potential value for MLC QA in highly dynamic treatment deliveries. Nonetheless, its pronounced energy and scanning-orientation dependence underscores the need for rigorous and consistent calibration and handling procedures.