April 13, 2026Journal of Clinical Laboratory Analysis0 citationsOpen Access

Lipid Interference Flagging and Haemoglobin Correction in Lipemic Samples With a Hematology Analyzer

View Full Paper
YFYinhua FanYSYucheng ShenKLK. Liu

Key Points

  • This research aims to evaluate the effectiveness of the lipid particles flag in hematology analyzers to enhance hemoglobin accuracy in lipemic samples.
  • Assessment of lipid interference flag in hematology analyzers
  • Comparison of HGB-O measurements in lipemic versus non-lipemic samples
  • Evaluation of accuracy in hemoglobin readings
  • The lipid particles flag effectively identifies lipid interference.
  • HGB-O provides reliable hemoglobin measurements in the presence of lipids.
  • Findings support enhanced accuracy in clinical testing.

Abstract

These findings highlight the clinical value of the "Lipid Particles?" flag in identifying lipid interference, demonstrating the accuracy of HGB-O in providing a reliable haemoglobin measurement.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Fan et al. (2026) studied this question.

synapsesocial.com/papers/69dc88f43afacbeac03eac40https://doi.org/10.1002/jcla.70223
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Lipemia: causes, interference mechanisms, detection and management2014 · 283 citations
  2. 2Intravenous Lipid Emulsions in Parenteral Nutrition2015 · 92 citations
  3. 3Development, History, and Future of Automated Cell Counters2015 · 105 citations
  4. 4Endogenous Interferences in Clinical Laboratory Tests2012 · 26 citations
  5. 5Effects of Insulin Resistance and Type 2 Diabetes on Lipoprotein Subclass Particle Size and Concentration Determined by Nuclear Magnetic Resonance2003 · 613 citations