August 29, 2008Journal of Neurosurgery Pediatrics77 citations

Magnetic field interactions in adjustable hydrocephalus shunts

View Full Paper
ALAndrea LavinioSHSally G. HardingFBFloor Van Der Boogaard

Key Points

Key points are not available for this paper at this time.

Abstract

All valves, with the exception of the Polaris and ProGAV models, are prone to unintentional reprogramming when exposed to heterogeneous magnetic fields stronger than 40 mT. All tested valves can be considered safe for 3-T MR imaging. All valves generated a distortion of the MR image, especially the GE sequences.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Lavinio et al. (2008) studied this question.

synapsesocial.com/papers/69debb1940ea065679559405https://doi.org/10.3171/ped/2008/2/9/222