Two-dimensional ultrafast transient absorption spectrograph covering deep-ultraviolet to visible spectral region optimized for biomolecules

View Full Paper
MNMaryam NazariNMNina Mosimann-SchönbächlerARAaron Riede

Key Points

Key points are not available for this paper at this time.

Abstract

Abstract We report on the implementation of a multi-kHz single-shot referenced non-coherent two-dimensional UV spectrograph based on conventional pump-probe geometry. It has the capability to cover a broad spectral region in excitation from 270-to-380 nm and in the detection from 270-to-390 nm and 320-to-720 nm. Other setups features are: an unprecedented time resolution of 33 fs (standard deviation); signals are photometrically corrected; a single-shot noise of <1 mOD. It has the capability to operate with sample volumes as small as few μ l which is an accomplishment in studying biological or biomimetic systems. To show its performances and potentials, we report two preliminary studies on the photophysics of phenanthrenes hosted in a multichromophoric antenna system and of aromatic amino acids in a blue-copper azurin.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Nazari et al. (2021) studied this question.

synapsesocial.com/papers/69edba5190efd21756be7c4fhttps://doi.org/10.1088/2515-7647/ac0805