Performance Evaluation of Instrument-Based SARS-CoV-2 Rapid Antigen Fluorescent Immunoassays for Point-of-Care Detection

View Full Paper
VKVidya KeshavLSLesley ScottLHLucia Hans

Key Points

  • This evaluation aims to assess the analytical and clinical performance of two rapid antigen tests for SARS-CoV-2.
  • Retrospective analytical evaluation of two fluorescent immunoassays (PCL and LumiraDx)
  • Assessment of clinical performance using 110 residual specimens, stratified by cycle threshold (Ct) against RT-PCR
  • Operational characteristics evaluated through a structured Likert framework.
  • PCL sensitivity was 63% (51–73), LumiraDx sensitivity was 95% (88–99) with overall specificity ≥97%.
  • Sensitivity increased significantly for Ct ≤ 25: 93% for PCL and 100% for LumiraDx.
  • LumiraDx maintained significant sensitivity (83–94%) at Ct 25–30, while PCL missed positives in this range.

Abstract

Rapid antigen tests targeting SARS-CoV-2 nucleocapsid protein were essential for decentralised testing during the COVID-19 pandemic. Independent performance evaluations are essential to support regulatory approval and inform clinical implementation, particularly in resource-limited settings. This study presents a retrospective analytical and operational evaluation of two instrument-based fluorescent immunoassays (FIAs): the PCL COVID-19 Ag Rapid FIA and LumiraDx SARS-CoV-2 Ag Test. Analytical sensitivity was determined using recombinant nucleocapsid protein and viral cultures. Clinical performance was assessed using residual clinical specimens (n = 110) with RT-PCR as a reference, stratified by cycle threshold (Ct). Operational characteristics were assessed using a structured Likert framework. Overall sensitivity was 63% (51–73) for PCL and 95% (88–99) for LumiraDx. For Ct ≤ 25, sensitivity increased to 93% and 100%. Specificity was ≥97% for both. LumiraDx maintained sensitivity (83–94%) at Ct 25–30, whereas PCL did not detect any positives in this range. The limit of detection was 39 pM (PCL) and 0.6 pM (LumiraDx). Operational usability was high for both (90% PCL, 87% LumiraDx). LumiraDx showed higher analytical sensitivity across a broader viral load range, supporting primary diagnostic use, whereas PCL was limited to high viral loads. This evaluation provides a reproducible framework for rapid diagnostic assessment during emerging outbreaks.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Keshav et al. (2026) studied this question.

synapsesocial.com/papers/69fadad703f892aec9b1e888https://doi.org/10.3390/covid6050078
Ask AI
Helpful
Bookmark
Share
View Full Paper