PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
June 9, 2010Journal of Cardiovascular Electrophysiology388 citations

Left Atrial Appendage Studied by Computed Tomography to Help Planning for Appendage Closure Device Placement

View Full Paper
YWYan WangLBLuigi Di BiaseRHRodney Horton

Structured PICO

What are the morphologic characteristics of the left atrial appendage as assessed by cardiac CT to aid in closure device placement?

P
Population
612 patients with or without atrial fibrillation (AF)
I
Intervention
64-channel cardiac computed tomography (CT)
O
Outcome
Morphologic parameters of the left atrial appendage (LAA) and LAA ostiumsurrogate

Cardiac CT reveals significant heterogeneity in LAA morphology and suggests that calculating LAA ostium diameter from its perimeter is the optimal method for sizing closure devices.

Abstract

OBJECTIVE: To quantitatively study various morphologic parameters of the left atrial appendage (LAA) by computed tomography (CT) to aid the preoperative planning and implantation of left atrial appendage closure devices. METHODS: In 612 cases of patients with or without atrial fibrillation (AF), a cardiac CT study was performed. RESULTS: The classification of general LAA morphology included ChickenWing type (18.3%), WindSock (46.7%), Cauliflower type (29.1%), and Cactus type (5.9%). Anatomical relationship of the LAA to the left superior pulmonary vein (LSPV) were classified as high type (superior to LSPV, 30.2%), mid type (parallel to LSPV, 58.1%), and low type (inferior to LSPV, 11.7%). LAA ostium could be classified into 5 types including oval (68.9%), foot-like (10%), triangular (7.7%), water drop-like (7.7%), and round (5.7%). Two-dimensional (2D) orthogonal method was obviously not accurate for determining the LAA orifice because the measurement was often unparallel to the LAA orifice. Two-dimensional oblique method was better than 3-dimensional method in reproducibility to determine the size of LAA ostium. The diameter calculated from the perimeter of the LAA ostium was superior to the diameter from direct measurement of the LAA ostium for selecting the occluder. CONCLUSION: The morphology of the LAA and the LA ostium are extremely complex and heterogeneous. Sixty-four-channel cardiac CT could assist preoperative planning of LAA closure device placement. The diameter of the LAA ostium calculated from the perimeter is the best parameter for sizing the LAA occluder.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Wang et al. (2010) studied this question.

synapsesocial.com/papers/69ff5e03da5c1eb07f2d7498https://doi.org/10.1111/j.1540-8167.2010.01814.x
Ask AI
Helpful
Bookmark
Share
View Full Paper