March 31, 2026HeartRhythm Case Reports1 citationsOpen Access

Septal Implantation of a Leadless Atrial Pacemaker: Relevant Anatomy for Alternate Sites into Bachmann’s Area

ABAlexandra BrigidoJMJake MartinezRTRoderick Tung

Structured PICO

P
Population
1 78-year-old woman with sick sinus syndrome, symptomatic bradycardia (heart rates in the 30s), non-perfusing ventricular escape beats, chronotropic incompetence, and preserved atrioventricular (AV) conduction.
I
Intervention
Leadless atrial pacemaker (Aveir AR) implanted in the interatrial septum (anterior to the fossa ovalis) guided by intracardiac echocardiography (ICE).
O
Outcome
Device electrical parameters (capture threshold, sensing amplitude, impedance) and P-wave duration.

Septal implantation of a leadless atrial pacemaker guided by intracardiac echocardiography is a feasible alternative when standard right atrial appendage positioning fails, potentially improving interatrial synchrony.

Limitations

  • Long-term follow-up is necessary to rule out aortic injury from potential erosion into the noncoronary sinus of Valsalva.