Migration and Misalignment of MyVal Balloon Expandable Valve during Transcatether Aortic Valve Implatation

LRLuka Rotkvić

Key Result

A dislocated MyVal balloon-expandable valve during TAVI was successfully managed by deploying it in the descending aorta and implanting a new prosthesis at the aortic position.

Study Design

Type

Case Report (n=1)

Multicenter

No

Structured PICO

P
Population
82-year-old patient with aortic stenosis
I
Intervention
Transcatheter aortic valve implantation (TAVI) using MyVal balloon-expandable valve
O
Outcome
Prosthesis displacement/migration and subsequent managementsafety

Highlights the management of TAVI prosthesis migration by deploying the dislocated valve in the descending aorta and implanting a new valve in the correct position.

Limitations

  • Exact reason of migration is not known
  • Single case report

Abstract

Introduction: Transcatheter aortic valve implantation (TAVI) is a treatment method for aortic stenosis with proven efficacy and recommended by professional societies in the cardiovascular field. Complications during TAVI are rare but important and potentially devastating, especially embolization or migration of prothesis. Case presentation: We present a case of an 82-year-old patient in whom the prosthesis was displaced during implantation and later expanded in the descending aorta with a new prosthesis implanted in the aortic valve position. Conclusion: Dislocation of the prosthesis during implantation is a serious complication. In most cases the prosthesis can be repositioned elsewhere in the aorta or perform a valve-in-valve procedure. In case of failure, the availability of cardiac surgery is important. Technological improvements, increasing operator experience and adequate periprocedural preparation minimize the risk of complications.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Luka Rotkvić (2024) conducted a case report in Severe aortic stenosis (n=1). Transcatheter aortic valve implantation (TAVI) with MyVal balloon expandable prosthesis was evaluated on Procedural success and management of complication. A dislocated MyVal balloon-expandable valve during TAVI was successfully managed by deploying it in the descending aorta and implanting a new prosthesis at the aortic position.