May 1, 2003IEEE Engineering in Medicine and Biology Magazine464 citations

Mobile monitoring with wearable photoplethysmographic biosensors

View Full Paper
HAH. Harry AsadaPSP. ShaltisARAndrew Reisner

Key Result

A novel wearable ring sensor using advanced photoplethysmographic techniques provides reliable ambulatory monitoring of heart rate, oxygen saturation, and heart rate variability.

Structured PICO

I
Intervention
Wearable biosensor (ring sensor) with photoplethysmographic (PPG) techniques

A novel wearable ring sensor using advanced photoplethysmographic techniques offers a promising ambulatory method for continuous monitoring of cardiovascular parameters.

Abstract

We address both technical and clinical issues of wearable biosensors (WBS). First, design concepts of a WBS are presented, with emphasis on the ring sensor developed by the author's group at MIT. The ring sensor is an ambulatory, telemetric, continuous health-monitoring device. This WBS combines miniaturized data acquisition features with advanced photoplethysmographic (PPG) techniques to acquire data related to the patient's cardiovascular state using a method that is far superior to existing fingertip PPG sensors. In particular, the ring sensor is capable of reliably monitoring a patient's heart rate, oxygen saturation, and heart rate variability. Technical issues, including motion artifact, interference with blood circulation, and battery power issues, are addressed, and effective engineering solutions to alleviate these problems are presented. Second, based on the ring sensor technology the clinical potentials of WBS monitoring are addressed.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Asada et al. (2003) conducted a review in Cardiovascular monitoring. Wearable photoplethysmographic biosensors (ring sensor) vs. Existing fingertip PPG sensors was evaluated. A novel wearable ring sensor using advanced photoplethysmographic techniques provides reliable ambulatory monitoring of heart rate, oxygen saturation, and heart rate variability.

synapsesocial.com/papers/6a0780a5beabdc12ba079a94https://doi.org/10.1109/memb.2003.1213624
Ask AI
Helpful
Bookmark
Share
View Full Paper