PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
January 1, 1990IEEE Transactions on Biomedical Engineering359 citations

Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses (neuronal application)

View Full Paper
TRTimothy L. RoseEIC LaboratoriesLRLois S. RobbleeVerizon (United States)

Key Points

Key points are not available for this paper at this time.

Abstract

The charge injection limits of a Pt electrode using 0.2 ms charge balanced, biphasic current pulses ranged from 50 to 150 microC/cm2 geometric if the potential excursions of the electrode are kept below those at which H2 or O2 is produced. These charge densities are three to ten times smaller than the currently accepted value based on earlier experiments in which the reversible surface reactions were fully utilized and the pulse widths were longer.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Rose et al. (1990) studied this question.

synapsesocial.com/papers/6a0822a1280cd4e998e8a901https://doi.org/10.1109/10.61038
Ask AI
Helpful
Bookmark
Share
View Full Paper