A Rapid, Simple, Sensitive Method for Measuring Fibrinolytic Split Products in Human Serum

View Full Paper
CMC. MerskeyPLParviz LalezariAJAlison Johnson

Key Points

  • This research aims to establish a sensitive method for detecting and quantifying fibrinolytic split products in human serum.
  • Utilizes sensitized human group O cells for detection.
  • Test can be performed using Microtiter and Autotiter systems, allowing for multiple samples.
  • Procedure requires only 1 hour for completion.
  • Test detects fibrinogen or split products at sensitivity of 1.0 μg/ml.
  • Fibrin split products measured accurately, while fibrinogen split products are less reliable due to absorption.
  • Test is particularly useful for diagnosing defibrination syndrome.

Abstract

Summary and ConclusionsA simplified procedure for recognizing and quantitating fibrinolytic split products has been described. It utilizes human group O cells which can be sensitized and stored at 4° for 3 weeks and at ‐20° for many months prior to use. It is very sensitive, detecting 1.0 μg/ml of fibrinogen or split products. Application of the test to the Microtiter and Autotiter systems permits assay of many samples at once. Only 1 hr is required to complete the test and read the results. The interpretation of the quantitative assay result should be carefully made since, while fibrin split products are not absorbed from the sera during the preparation of the sample and are relatively accurately measured, fibrinogen split products tend to be absorbed during sample preparation and are thus less accurately measured. The test would appear to be more useful with fibrin lysis than with fibrinogen lysis and to be especially applicable in the diagnosis of defibrination syndrome.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Merskey et al. (1969) studied this question.

synapsesocial.com/papers/6a090abf73218fa1919d21ffhttps://doi.org/10.3181/00379727-131-33998
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1A Rapid Method for the Semiquantitative Determination of Fibrinogen and Fibrinogen Degradation Products (FDP) in Defibrination Syndrome1971 · 3 citations
  2. 2A-144 Liquid chromatography multiple reaction monitoring-based assay for quantitative assessment of fibrinogen levels in clinical samples2024
  3. 3Fibrin fragment D-dimer and fibrinogen B beta peptides in plasma as markers of clot lysis during thrombolytic therapy in acute myocardial infarction1990 · 38 citations
  4. 4Methods, precision, and analytical sensitivity of a novel low‐plasma‐volume assay of fibrinolytic capacity utilizing the euglobulin fraction2024 · 2 citations
  5. 5Rapid Fibrinogen Assessment With the qLabs FIB Point‐of‐Care System: Results From a Multicenter Clinical Validation Study2026