May 17, 2026Journal of Ophthalmology0 citationsOpen Access

Evaluation of a New Algorithm‐Based Approach in Subjective Refraction

View Full Paper
ISImène Salah-MabedVPVincent PeanDGDamien Gatinel

Key Points

  • This study evaluates the efficacy of an algorithm-based approach in subjective refraction by comparing it with expert manual methods.
  • Prospective comparison of manual refraction by an expert and automated refraction using the SiviewExam Expert system.
  • Conducted at Rothschild Foundation from October 2022 to May 2023 with 107 patients.
  • Outcomes analyzed using power vector analysis for three dioptric components: M, J0, and J45.
  • Mean differences showed no bias in measurements for M (0.01 D) and J45 (0.009 D), with a clinically negligible bias for J0 (0.05 D).
  • Limits of agreement were significantly lower than inter-examiner variability (0.7 D).
  • The SiviewExam Expert system report was 100% accurate compared to expert evaluations.

Abstract

Purpose The aim of this study was to evaluate a new algorithm‐based approach in subjective refraction by comparing the refractions obtained by a professional expert with those obtained using the SiviewExam Expert system on a large population. Methods Two subjective refraction methods were compared prospectively on patients visiting Rothschild Foundation between October 2022 and May 2023: (i) A manual refraction performed by an expert using a standardized method and (ii) an automated one with the SiviewExam Expert system. Outcomes of the two subjective refractions were compared using a power vector analysis method. Power vectors are a geometric representation of spherocylindrical refractive errors in 3 fundamental dioptric components: M ( M = S + C /2), J0 (J0 = (− C /2) cos (2 α ) at an axis of α = 0 = 180°), and J45 (J45 = (− C /2) sin (2 α ) at an axis of α = 45°). At the end of the examination, the SiviewExam Expert system provided a report that has been analyzed as well. Results A total of 107 patients, with a mean age of 35.7 ± 11.9 years (ranging from 19 to 69 years) were included in the study. The mean SE was −2.10 ± 3.37 D (ranging from −11 D to +7.13 D). The mean differences have shown that there were no bias between the measurements for M (0.01 D, −0.04–0.07 D) and J45 (0.009 D, −0.01–0.03 D) and a clinically negligible bias (0.05 D, −0.08–−0.02 D) in the estimation of J0. The limits of agreement were −0.54 D (lower limit: −0.63–−0.44) to 0.56 D (upper limit: 0.47–0.65), −0.22 D (lower limit: −0.26 to −0.18) to 0.24 D (upper limit: 0.20–0.27), –0.34 D (lower limit: –0.39 to –0.29) to 0.25 D (upper limit: 0.20 to 0.30) for M , J45, and J0, respectively, and were significantly lower than the expected inter‐examiner variability of 0.7 D. Finally, the report provided by the SiviewExam Expert system was found to be accurate according to the expert in 100% of the cases. Conclusion Our outcomes show that in a large adult population of over 100 patients with a wide range of ametropias, there were no statistically significant differences in the results for M , J0, and J45 between the two refraction methods, indicating that we could reasonably clinically consider the SiviewExam Expert system examination comparable with the one conducted by an expert.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Salah-Mabed et al. (2026) studied this question.

synapsesocial.com/papers/6a095c5d7880e6d24efe280dhttps://doi.org/10.1155/joph/5572620
Ask AI
Helpful
Bookmark
Share
View Full Paper