October 19, 2021Scientific Reports55 citationsOpen Access

Validation of the optical Aktiia bracelet in different body positions for the persistent monitoring of blood pressure

JSJosep SolàAVAnna VybornovaSFSibylle Fallet

Key Result

The Aktiia Bracelet estimated systolic BP compared to auscultation with mean differences of 0.46 mmHg sitting, -2.44 mmHg lying, and -0.62 mmHg standing over 4 weeks.

Study Design

Type

Observational (n=91)

Structured PICO

Does a wearable optical blood pressure monitor (Aktiia Bracelet) accurately estimate blood pressure in different body positions compared to manual double-auscultation in adults?

P
Population
91 male and female adults aged 21-65 years, with blood pressure ranging from hypotensive to hypertensive levels. Excluded: wrist circumference > 22 cm.
I
Intervention
Aktiia Bracelet (wearable optical blood pressure monitor) worn on the right wrist for persistent monitoring
C
Comparator
Manual double-blinded auscultatory blood pressure measurements (sphygmomanometer on left upper arm) with a volume-clamp device (Nexfin) on the left middle finger as a reference-control
O
Outcome
Mean and standard deviation of the differences of the readings of the two measurement modalities (Aktiia vs auscultation) for systolic and diastolic blood pressure in each body positionsurrogate

The Aktiia wearable optical bracelet provides reasonably accurate blood pressure estimates across different body positions compared to auscultation, demonstrating the feasibility of persistent, cuffless blood pressure monitoring.

Limitations

  • Precision varies across body positions
  • Prognostic value for cardiovascular events and association with target organ damage requires further cross-sectional and longitudinal studies
  • Acceptance rates for generating readings were lower than 70% for some body positions (e.g., 45% in standing position)
  • Potential position-dependent calibration bias since the device was initialized only in the sitting position
  • Inherent uncertainty in reference auscultatory measurements when performed in non-sitting positions

Abstract

The diagnosis of hypertension and the adjustment of antihypertensive drugs are evolving from isolated measurements performed at the physician offices to the full phenotyping of patients in real-life conditions. Indeed, the strongest predictor of cardiovascular risk comes from night measurements. The aim of this study was to demonstrate that a wearable device (the Aktiia Bracelet) can accurately estimate BP in the most common body positions of daily life and thus become a candidate solution for the BP phenotyping of patients. We recruited 91 patients with BP ranging from low to hypertensive levels and compared BP values from the Aktiia Bracelet against auscultatory reference values for 4 weeks according to an extended ISO 81060-2 protocol. After initializing on day one, the observed means and standard deviations of differences for systolic BP were of 0.46 ± 7.75 mmHg in the sitting position, - 2.44 ± 10.15 mmHg in the lying, - 3.02 ± 6.10 mmHg in the sitting with the device on the lap, and - 0.62 ± 12.51 mmHg in the standing position. Differences for diastolic BP readings were respectively of 0.39 ± 6.86 mmHg, - 1.93 ± 7.65 mmHg, - 4.22 ± 6.56 mmHg and - 4.85 ± 9.11 mmHg. This study demonstrates that a wearable device can accurately estimate BP in the most common body positions compared to auscultation, although precision varies across positions. While wearable persistent BP monitors have the potential to facilitate the identification of individual BP phenotypes at scale, their prognostic value for cardiovascular events and its association with target organ damage will need cross-sectional and longitudinal studies. Deploying this technology at a community level may be also useful to drive public health interventions against the epidemy of hypertension.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Solà et al. (2021) conducted an observational in Hypertension (n=91). Aktiia Bracelet vs. Auscultatory reference values was evaluated on Mean and standard deviation of differences for systolic and diastolic BP in various body positions. The Aktiia Bracelet estimated systolic BP compared to auscultation with mean differences of 0.46 mmHg sitting, -2.44 mmHg lying, and -0.62 mmHg standing over 4 weeks.