Mechanisms of Residual Lumen Stenosis After High-Pressure Stent Implantation

View Full Paper
JBJavier BermejoJBJavier BotasEGEulogio Garcı́a

Key Result

High-pressure stent implantation achieves only 57% of maximal achievable lumen dimensions, primarily due to balloon subexpansion (33%) and elastic recoil (20%) (both P<0.0001).

Study Design

Type

Observational (n=50)

Structured PICO

P
Population
50 patients with 57 lesions (44 de novo, 13 restenotic) undergoing elective high-pressure stenting
I
Intervention
High-pressure stent implantation (median 14+/-2 atm) using Wiktor or Palmaz-Schatz stents with intravascular ultrasound (IVUS) assessment
O
Outcome
Mechanisms of residual lumen stenosis (balloon subexpansion and elastic recoil)surrogate

Despite high-pressure deployment, final stent lumen dimensions are only 57% of the maximal achievable due to inadequate balloon expansion and elastic recoil.

Main Result

p-value: p=<0.0001

Abstract

BACKGROUND: Intravascular ultrasound (IVUS) studies have demonstrated that stents are frequently suboptimally expanded despite the use of high pressures for deployment. The purpose of this study was to identify the mechanisms responsible for such residual lumen stenosis. METHODS AND RESULTS: Fifty-seven lesions from 50 patients treated with high-pressure (median+/-interquartile range, 14+/-2 atm) elective (44 de novo, 13 restenotic lesions) stenting were prospectively studied (29 Wiktor, Medtronic; 28 Palmaz-Schatz, Cordis Corp). Balloon subexpansion was calculated as the difference between maximal and minimal balloon cross-sectional areas at peak pressure measured by automatic edge detection; elastic recoil was calculated as the difference between minimal measured balloon size and IVUS-derived minimal lumen area within the stent. Angiographic residual diameter stenosis was 10+/-13% (reference diameter, 3.1+/-0.7 mm; balloon to artery ratio, 1.12+/-0.23) and IVUS-derived stent expansion was 80+/-28%. However, although balloon nominal size was 9.6+/-1.3 mm2 and maximal balloon size measured inside the coronary lumen was 12.5+/-3.2 mm2, final stent minimal lumen area was only 7.1+/-2.2 mm2. Balloon subexpansion of 4.0+/-1.8 mm2 (33%) and elastic recoil of 1.6+/-2.3 mm2 (20%) (both P<0.0001) were the two mechanisms responsible for residual luminal stenosis. Wiktor stent and peak inflation pressure correlated with balloon subexpansion, whereas Wiktor stent, de novo lesion, and minimal lumen area at baseline correlated with elastic recoil. CONCLUSIONS: Despite high-pressure deployment, lumen dimensions after stenting are only 57% of maximal achievable. Inadequate balloon expansion and elastic recoil are responsible for residual lumen stenosis, suggesting that plaque characteristics and stent resistance deserve further investigation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Bermejo et al. (1998) conducted an observational in Coronary artery stenosis (n=50). High-pressure stent implantation was evaluated on Mechanisms of residual lumen stenosis (balloon subexpansion and elastic recoil) (p=<0.0001). High-pressure stent implantation achieves only 57% of maximal achievable lumen dimensions, primarily due to balloon subexpansion (33%) and elastic recoil (20%) (both P<0.0001).