May 20, 2026American Journal of Respiratory and Critical Care Medicine0 citations

D92-06 Airway Scaffolds Improve Ventilation Heterogeneity in Patients With Emphysema-related Hyperinflation

View Full Paper
ATA TanaAKA KarirLCL T Chan

Key Points

  • To assess changes in ventilation heterogeneity following Apreo BREATHE Scaffold treatment in patients with emphysema-related hyperinflation.
  • Twenty-one subjects with emphysema were treated bilaterally with three scaffolds per lung.
  • Inspiratory and expiratory non-contrast CT scans were obtained at baseline and six months post-treatment.
  • Ventilation heterogeneity was quantified using CT:VQ technology, and paired analysis was used to compare lung volumes and VH.
  • End-inspiratory volume decreased by 184 mL (p = 0.0008) and end-expiratory volume by 593 mL (p = 0.004).
  • Mean ventilation heterogeneity decreased by 9.8% overall (p = 0.19).
  • In patients with baseline VH score ≥ 50 (n = 12), VH decreased by 20.3% (p = 0.012), indicating significant improvement in ventilation distribution.

Abstract

Abstract Rationale Ventilation heterogeneity (VH) refers to uneven regional ventilation, leading to ventilation–perfusion (V/Q) mismatch, inefficient gas exchange, and impaired respiratory mechanics. In emphysema, VH may contribute to dyspnoea and exercise limitation. CT:VQ by 4DMedical is an FDA-approved imaging technology that converts standard non-contrast chest CT scans into quantitative ventilation (V) and perfusion (Q) maps, enabling objective assessment of VH. The Apreo BREATHE Airway Scaffold is a self-expanding nitinol implant deployed bronchoscopically to release trapped air from hyperinflated emphysematous regions and improve lung mechanics. While prior studies have demonstrated significant improvements in symptoms, exercise capacity, and lung function, the impact of this device on VH has not been established. The aim was to evaluate changes in VH following Apreo BREATHE Scaffold treatment using CT:VQ imaging. Methods Twenty-one subjects with emphysema-related hyperinflation were treated bilaterally with three scaffolds per lung at the Royal Brompton Hospital as part of the first-in-human BREATHE-2 study. Inspiratory and expiratory non-contrast CT scans were acquired at baseline and six months post-treatment. VH was quantified using CT:VQ technology. A subgroup analysis included patients with a baseline VH score ≥ 50. Changes in end-inspiratory volume (EIV), end-expiratory volume (EEV), and VH were compared using paired statistical analysis. Results Eighteen subjects had paired scans meeting CT:VQ quality criteria. The mean age was 62.9 years, mean smoking exposure was 40.5 pack-years, and 21% were male. Following treatment, EIV and EEV decreased by 184 mL (p = 0.0008) and 593 mL (p = 0.004), respectively. Mean VH decreased by 9.8% (p = 0.19) overall. Among patients with a baseline VH score ≥ 50 (n = 12), VH decreased by 20.3% (p = 0.012), indicating significantly improved ventilation distribution in those with the highest initial heterogeneity. Conclusion Apreo BREATHE Scaffold treatment was associated with reductions in lung volumes and improved ventilation distribution, as demonstrated by CT:VQ imaging. The greatest VH improvement occurred in patients with high baseline heterogeneity, suggesting that VH may serve as a physiological biomarker for treatment response. Larger controlled studies are warranted to confirm these findings and further define the clinical utility of VH in assessing emphysema therapies and V/Q mismatch correction. This abstract is funded by: Data derived from the study was funded by Apreo Health

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Tana et al. (2026) studied this question.

synapsesocial.com/papers/6a0d5025f03e14405aa9bd33https://doi.org/10.1093/ajrccm/aamag162.3611
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1C23-04 The Ventilation-Perfusion Discrepancy Index Assists Target Lobe Selection for Endobronchial Valve Lung Volume Reduction: A Prospective Study2026
  2. 2C23-01 Enhancing Bronchoscopic Lung Volume Reduction Outcomes Through Advanced Imaging of Collateral Ventilation2026
  3. 3Regional ventilation–perfusion changes after endobronchial valve therapy assessed by single-photon emission computed tomography in adults with severe COPD2026
  4. 4C73-14 Defects From Novel Contrast-free CT 3d Ventilation- Perfusion Mismatch Imaging Are Related to Measurements of Gas Transfer2026
  5. 5Dynamic Imaging of Physiological Dead Space and Alveolar Ventilation in Patients Undergoing Endobronchial Valve Therapy with HP Xenon MRI2025