January 1, 1991Radiocarbon17 citationsOpen Access

Lithium Contamination in AMS Measurements of 14C

DLD. H. LoydJVJohn S. VogelSTSusan Trumbore

Key Points

Key points are not available for this paper at this time.

Abstract

High count rates of lithium ions in an AMS measurement system for 14 C were traced to the lithium content in the quartz tubes used in the graphitization of the samples. The lithium contamination was nearly eliminated by the use of borosilicate glass reaction tubes at a lower reaction temperature. The ion beam current and the measurement precision of the isotope ratio were not affected.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Loyd et al. (1991) studied this question.

synapsesocial.com/papers/6a0df719cae7912d2fa5708bhttps://doi.org/10.1017/s0033822200040327