PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
September 1, 2001Heart115 citationsOpen Access

Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators

MNMichael NiehausStadtarchiv GifhornJTJürgen TebbenjohannsUniversity of Hildesheim

Key Result

Electromagnetic interference from common sources like household appliances and cellular phones kept at least 10 cm away poses little hazard to patients with implanted pacemakers or ICDs.

Structured PICO

Does exposure to common sources of electromagnetic interference cause clinically significant malfunction in implanted pacemakers or ICDs?

P
Population
Patients with implanted pacemakers or implantable cardioverter-defibrillators (ICDs)
I
Intervention
Exposure to sources of electromagnetic interference (EMI) including microwaves, metal detectors, cellular phones, electronic article surveillance, MRI, electrocautery, and radiotherapy
O
Outcome
Device malfunction or electromagnetic interference (e.g., oversensing, inhibition, inappropriate shocks, resetting)safety

Common environmental electromagnetic sources pose minimal risk to modern pacemakers and ICDs if basic precautions, such as keeping mobile phones 10 cm away, are followed.

Abstract

Electromagnetic radiation may interact adversely with implanted pacing systems and implantable cardioverter-defibrillators (ICDs).1-4 Modern life exposes us all to an ever increasing number of potential sources of electromagnetic interference (EMI) and patients with implanted pacemakers or ICDs often ask about the use of microwave ovens, walking through airport metal detectors, and the use of cellular phones. This article gives an overview of the current knowledge of the effects of EMI on pacemakers and ICDs. Although no recent studies have been performed which test the effect of household microwave energy on pacemakers and ICDs, it is widely believed and accepted that all modern pacemakers are adequately shielded from microwave energy produced by modern appliances.5 Pacemaker manufacturers therefore recommend that patients with implanted devices do not need to take special precautions in the use of microwaves or other common household equipment such as televisions, radios, toasters, and electric blankets. The effect of metal detector gates on implanted pacemakers was studied more than 10 years ago.6 In 103 patients who were monitored as they passed through typical metal detectors, alarms invariably were activated when the patients walked through the gates. In none of the patients was the pacemaker function affected. None of the devices was reset to the programmed noise protection mode (most often V00) or spontaneous fixed rate mode of function, nor were any of the devices' outputs inhibited in paced patients, or inappropriately delivered in patients who had normal cardiac rhythm. It is therefore accepted practice to advise patients that while airport screening devices may detect the pacemaker or ICD metal case the device will not be adversely affected. Patients should carry their device identification card for the purpose of obtaining security clearance. Several studies have shown that cellular phones might cause EMI with complex medical equipment7 …

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Niehaus et al. (2001) conducted an editorial in Implanted pacemakers or cardioverter-defibrillators. Electromagnetic interference (EMI) was evaluated. Electromagnetic interference from common sources like household appliances and cellular phones kept at least 10 cm away poses little hazard to patients with implanted pacemakers or ICDs.

synapsesocial.com/papers/6a0e9bbb1c7919c41b4c3a00https://doi.org/10.1136/heart.86.3.246
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1The Effect of Metal Detector Gates on Implanted Permanent Pacemakers1988 · 57 citations
  2. 2Evaluation of Effects of the Microwave Oven (915 and 2450 MHz) and Radar (2810 and 3050 MHz) Electromagnetic Radiation on Noncompetitive Cardiac Pacemakers1973 · 19 citations
  3. 3Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-Defibrillators1999 · 71 citations
  4. 4Cellular Phone Interference Testing of Implantable Cardiac Defibrillators In Vitro1998 · 58 citations
  5. 5Electromagnetic Interference of Digital and Analog Cellular Telephones with Implantable Cardioverter Defibrillators: In Vitro and In Vivo Studies1999 · 47 citations