January 1, 1999Blood Pressure Monitoring22 citations

Evaluation of the Schiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation

View Full Paper
EOEoin O’BrienFMFáinsía MeeNANeil Atkins

Key Result

The Schiller BR-102 monitor achieved BHS grade B and satisfied AAMI accuracy criteria for blood pressure in auscultatory mode, but failed to provide accurate measurements in oscillometric mode.

Study Design

Type

Observational (n=85)

Structured PICO

Does the Schiller BR-102 ambulatory blood pressure monitor provide accurate measurements compared to standard criteria in subjects with a wide range of blood pressures?

P
Population
85 subjects with a wide range of blood pressures for the main validation test (plus 30 subjects for the in-use field phase)
I
Intervention
Schiller BR-102 ambulatory blood pressure monitor (evaluated in both auscultatory and oscillometric modes)
C
Comparator
Standard mercury sphygmomanometer measurements
O
Outcome
Accuracy of blood pressure measurement according to the British Hypertension Society (BHS) grading system and the Association for the Advancement of Medical Instrumentation (AAMI) standardsurrogate

The Schiller BR-102 ambulatory blood pressure monitor provides accurate measurements in auscultatory mode, but its oscillometric mode does not meet standard accuracy criteria.

Abstract

OBJECTIVE: To evaluate the Schiller BR-102 monitor for ambulatory blood pressure measurement according to the protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI). DESIGN: The BHS protocol is divided into two parts. Part I, which is the part applicable to this study, comprises the main validation procedure and has five phases: before-use device calibration; in-use (field) phase; after-use device calibration; static device validation; and report of evaluation. METHOD: Three Schiller BR-102 recorders passed the before-use device calibration test, after which they entered the in-use (field) assessment phase during which the three recorders were each worn by 10 subjects for 24 h, after which calibration was again assessed. Because there was no difference in results of calibration testing among the three devices, one was selected randomly and the main validation test was carried out on 85 subjects with a wide range of blood pressures both for the auscultatory mode and for the oscillometric mode using the Sphygmocorder. The results were analysed according to the BHS grading system from A to D. The data were also analysed according to the standard of the Association for the Advancement of Medical Instrumentation (AAMI), which stipulates that the mean difference between the test device and the standard shall be 160/100 mmHg) the Schiller BR-102 was less accurate in the high pressure range for diastolic blood pressure but more accurate for systolic blood pressure, achieving A/C grading, while satisfying the AAMI criteria both for systolic and for diastolic blood pressure in the auscultatory mode. In the oscillometric mode the device performed less accurately in the high-pressure range, achieving grade D/C, while failing to satisfy the AAMI criteria both for systolic and for diastolic blood pressure. The means+/-SD of the first mercury sphygmomanometer measurements were 143+/-32 mmHg for systolic blood pressure and 88+/-21 mmHg for diastolic blood pressure. Acceptability to subjects was good and the manufacturer's manual was satisfactory. CONCLUSION: On the basis of these results, the Schiller BR-102 can be recommended for ambulatory blood pressure measurement in clinical practice using the auscultatory mode, but the oscillometric mode, which operates only if the device fails in the auscultatory mode, does not provide accurate measurements.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

O’Brien et al. (1999) conducted an observational in Blood pressure measurement (n=85). Schiller BR-102 ambulatory blood pressure system vs. Standard mercury sphygmomanometer was evaluated on Accuracy of blood pressure measurement according to BHS and AAMI protocols. The Schiller BR-102 monitor achieved BHS grade B and satisfied AAMI accuracy criteria for blood pressure in auscultatory mode, but failed to provide accurate measurements in oscillometric mode.

synapsesocial.com/papers/6a0ec608b7cc3b883f22b98chttps://doi.org/10.1097/00126097-199904010-00006