PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
April 1, 2011Annals of Noninvasive Electrocardiology252 citationsOpen Access

Reliability of Ultra‐Short ECG Indices for Heart Rate Variability

View Full Paper
UNUdi NussinovitchKEKeren ElishkevitzKKKeren Katz

Key Result

Ultra-short 1-minute and 10-second ECG recordings showed good correlation with standard 5-minute recordings for average RR interval and RMSSD, but not for SDNN.

Key Points

  • This study aims to assess the reliability of heart rate variability (HRV) parameters from ultra-short electrocardiogram recordings.
  • Evaluated HRV for 5 minutes in 70 healthy volunteers.
  • Calculated HRV parameters from 1-minute and 10-second ECG recordings.
  • Correlated measurements using intraclass correlation coefficients.
  • Good correlations observed between 5-minute and both 1-minute (ICC not specified) and 10-second ECGs for average RR interval and RMSSD.
  • No correlation found for SDNN and other HRV parameters.
  • RMSSD is a reliable parameter for ultra-short recordings, while longer recordings are needed for comprehensive analysis.

Study Design

Type

Observational (n=70)

Structured PICO

Does ultra-short ECG recording (1-minute or 10-second) reliably measure heart rate variability parameters compared to standard 5-minute recordings in healthy volunteers?

P
Population
70 healthy volunteers
I
Intervention
Ultra-short electrocardiogram recordings (1-minute and 10-second intervals)
C
Comparator
Standard 5-minute electrocardiogram recordings
O
Outcome
Reliability of HRV parameters (correlation using intraclass correlation coefficients)surrogate

Ultra-short ECG recordings of 10 seconds or 1 minute can reliably assess RMSSD for heart rate variability, but not SDNN.

Abstract

BACKGROUND: Heart rate variability (HRV) is an accepted and reliable means for assessing autonomic nervous system dysfunction. A 5-minute measurement of HRV is considered methodologically adequate. Several studies have attempted to use shorter recordings of 1-2 minutes or 10 seconds. The aim of this study was to determine the reliability of HRV parameters calculated from ultra-short electrocardiogram recordings. METHODS: Seventy healthy volunteers were recruited for the study. HRV was evaluated for 5 minutes according to accepted procedures. Thereafter, HRV parameters were recalculated from randomly selected 1-minute and 10-second intervals. The standard and ultra-short measurements were correlated using intraclass correlation coefficients. RESULTS: Good correlations between the 5-minute electrocardiograms (ECGs) and both the 1-minute and 10-second ECGs were noted for average RR interval, and root mean square of successive differences in RR intervals (RMSSD). No correlation was noted for standard deviation of the RR interval (SDNN) and several other HRV parameters. CONCLUSIONS: RMSSD, but not SDNN, seem a reliable parameter for assessing HRV from ultra-short (1 minute or 10 seconds) resting electrocardiographic recordings. Power spectral analysis and evaluation of other HRV parameters require longer recording periods. Further research is required to evaluate the importance of ultra-short RMSSD for cardiovascular risk stratification.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Nussinovitch et al. (2011) conducted an observational in Healthy volunteers (n=70). Ultra-short ECG recordings (1-minute and 10-second) vs. Standard 5-minute ECG recordings was evaluated on Reliability of HRV parameters (intraclass correlation coefficients between standard and ultra-short measurements). Ultra-short 1-minute and 10-second ECG recordings showed good correlation with standard 5-minute recordings for average RR interval and RMSSD, but not for SDNN.

synapsesocial.com/papers/6a0eeef3218372ada647d958https://doi.org/10.1111/j.1542-474x.2011.00417.x
Ask AI
Helpful
Bookmark
Share
View Full Paper