October 29, 2007Journal of Magnetic Resonance Imaging39 citationsOpen Access

Real‐time MR imaging of myocardial regional function using strain‐encoding (SENC) with tissue through‐plane motion tracking

View Full Paper
EIEl‐Sayed H. IbrahimMSMatthias StuberAFAhmed S. Fahmy

Key Result

Real-time myocardial strain-encoding (SENC) imaging with slice-following yielded significantly different strain values compared to SENC without slice-following during systole (P<0.01).

Structured PICO

P
Population
10 human volunteers and 5 pigs (2 with induced myocardial infarction)
I
Intervention
Real-time myocardial strain-encoding (SENC) imaging with slice-following technique (3D selective excitation)
C
Comparator
SENC imaging without slice-following, orthogonal SPAMM-tagged images, and standard delayed-enhancement (DE) images
O
Outcome
Strain values and curvessurrogate

Correction of through-plane motion in real-time cardiac functional imaging using slice-following is feasible and provides more accurate strain measurements than conventional SENC.

Main Result

p-value: p=<0.01

Abstract

PURPOSE: To implement real-time myocardial strain-encoding (SENC) imaging in combination with tracking the tissue displacement in the through-plane direction. MATERIALS AND METHODS: SENC imaging was combined with the slice-following technique by implementing three-dimensional (3D) selective excitation. Certain adjustments were implemented to reduce scan time to one heartbeat. A total of 10 volunteers and five pigs were scanned on a 3T MRI scanner. Spatial modulation of magnetization (SPAMM)-tagged images were acquired on planes orthogonal to the SENC planes for comparison. Myocardial infarction (MI) was induced in two pigs and the resulting SENC images were compared to standard delayed-enhancement (DE) images. RESULTS: The strain values computed from SENC imaging with slice-following showed significant difference from those acquired without slice-following, especially during systole (P < 0.01). The strain curves computed from the SENC images with and without slice-following were similar to those computed from the orthogonal SPAMM images, with and without, respectively, tracking the tag line displacement in the strain direction. The resulting SENC images showed good agreement with the DE images in identifying MI in infarcted pigs. CONCLUSION: Correction of through-plane motion in real-time cardiac functional imaging is feasible using slice-following. The strain measurements are more accurate than conventional SENC measurements in humans and animals, as validated with conventional MRI tagging.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Ibrahim et al. (2007) studied Healthy volunteers and animal models (myocardial infarction) (n=15). Real-time myocardial strain-encoding (SENC) imaging with slice-following vs. SENC without slice-following and SPAMM-tagged images was evaluated on Strain values during systole (p=<0.01). Real-time myocardial strain-encoding (SENC) imaging with slice-following yielded significantly different strain values compared to SENC without slice-following during systole (P<0.01).

synapsesocial.com/papers/6a0f6ed09e54838161fcb6bfhttps://doi.org/10.1002/jmri.21125
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Human heart: tagging with MR imaging--a method for noninvasive assessment of myocardial motion.1988 · 1,428 citations
  2. 2Inducibility of Sustained Ventricular Tachycardia in a Closed‐Chest Ovine Model of Myocardial Infarction1999 · 24 citations
  3. 3Real‐time imaging of regional myocardial function using fast‐SENC2006 · 101 citations
  4. 4Real‐time imaging of two‐dimensional cardiac strain using a harmonic phase magnetic resonance imaging (HARP‐MRI) pulse sequence2003 · 76 citations
  5. 5Correction of through‐plane deformation artifacts in stimulated echo acquisition mode cardiac imaging2006 · 20 citations