February 7, 2005Journal of Cardiovascular Electrophysiology58 citations

Saline‐Irrigated Radiofrequency Ablation Electrode with External Cooling

View Full Paper
FWFred H.M. WittkampfHNHiroshi NakagawaSFSarah Foresti

Structured PICO

Does thermal insulation of irrigation channels in radiofrequency ablation electrodes improve temperature feedback and lesion controllability in a dog thigh muscle model?

P
Population
Thigh muscle preparation in dogs (5 thigh muscles of 3 dogs)
I
Intervention
Modified open flush, irrigated catheters (thermally insulating the irrigation channels inside the ablation electrode), 60 s, 20-50 Watt applications, constant saline flush rate of 20 cc/min
C
Comparator
Standard open flush, irrigated catheters (Sprinklr, Medtronic Inc) with same settings
O
Outcome
Lesion dimensions and maximum electrode temperature risesurrogate

Thermal insulation of irrigation channels in RF ablation electrodes improves temperature feedback and correlation with lesion volume without altering lesion dimensions.

Abstract

INTRODUCTION: Open flush, irrigated ablation electrodes may improve the safety of radiofrequency catheter ablation by preventing protein aggregation and coagulum formation. This is particularly important in left-sided procedures like catheter ablation of atrial fibrillation. Electrode cooling and the inherent loss of temperature feedback, however, grossly reduce the ability to monitor tissue heating. Intimate contact may not be recognized and the delivery of nominal RF power levels may then lead to excessive tissue heating, steam explosions, and even tamponade. METHODS AND RESULTS: Standard, open flush, irrigated catheters (Sprinklr, Medtronic Inc, Minneapolis, MN) were modified by thermally insulating the irrigation channels inside the ablation electrode. Using the thigh muscle preparation, multiple lesions were created with standard and modified catheters using 60 s, 20-50 Watt applications and a constant saline flush rate of 20 cc/min. A total of 57 lesions were created on five thigh muscles of three dogs. Lesion dimensions were not significantly different between both types of catheter, but the maximum electrode temperature rise during ablation was significantly higher with the modified catheter. Insulation of the irrigation channels improved the correlation coefficient between maximum electrode temperature rise and lesion volume from 0.38 (ns) to 0.62 (P < 0.001). CONCLUSION: Thermal insulation of the irrigation channels facilitates temperature feedback during radiofrequency ablation and controllability of lesion formation.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Wittkampf et al. (2005) studied this question.

synapsesocial.com/papers/6a0ff0d79e54838161fd70aehttps://doi.org/10.1046/j.1540-8167.2005.40629.x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Radiofrequency Catheter Ablation Using Cooled Electrodes: Impact of Irrigation Flow Rate and Catheter Contact Pressure on Lesion Dimensions2002 · 102 citations
  2. 2Temperature monitoring during radiofrequency catheter ablation of accessory pathways.1992 · 180 citations
  3. 3Soft Thrombus Formation in Radiofrequency Catheter Ablation2002 · 78 citations
  4. 4Thromboembolic Complications of Cardiac Radiofrequency Catheter Ablation1999 · 185 citations
  5. 5Irrigated‐Tip Catheter Ablation of Pulmonary Veins for Treatment of Atrial Fibrillation2002 · 152 citations