A simple method for obtaining highly accurate Pb isotope data by MC-ICP-MS

View Full Paper
JWJon Woodhead

Key Points

Key points are not available for this paper at this time.

Abstract

Many recent studies of isotope ratio determination by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) have highlighted the excellent levels of analytical precision obtainable using this newly emergent technology. Unfortunately, it is often a considerably more difficult matter to obtain highly accurate analyses by MC-ICP-MS. This is particularly the case for isotope systems in which internal normalisation for instrumental mass bias effects is not possible, e.g., Pb, Cu, Zn, B, Li. Given the fact that extremely accurate Pb isotope data can now be obtained by double-spike thermal ionisation mass spectrometry (TIMS), this is seemingly a major impediment to the widespread adoption of MC-ICP-MS methods for the analysis of Pb. It is demonstrated in this contribution, however, that a few simple modifications to existing protocols can produce MC-ICP-MS Pb isotope data of comparable accuracy to the best double spike TIMS results. This method also preserves many of the inherent advantages of ICP-MS analysis, such as rapid sample processing, and avoids many of the stringent analytical requirements of double spiking. The method is highly reliable and a summary of results from our laboratory over the past two years is presented, together with suggestions for further development.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jon Woodhead (2002) studied this question.

synapsesocial.com/papers/6a0ff29bd13714ec96fee095https://doi.org/10.1039/b205045e
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Multicollector ICP-MS analysis of Pb isotopes using a 207pb-204pb double spike demonstrates up to 400 ppm/amu systematic errors in Tl-normalization2002 · 322 citations
  2. 2Fractionation corrections for mixed-isotope spikes of Sr, K, and Pb1971 · 76 citations
  3. 3Lead isotopic analyses of NIST Standard Reference Materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effect1996 · 199 citations
  4. 4Pb, Sr and 10Be isotopic studies of volcanic rocks from the Northern Mariana Islands. Implications for magma genesis and crustal recycling in the Western Pacific1985 · 111 citations
  5. 5Investigation of matrix effects for Pb isotope ratio measurements by multiple collector ICP-MS: verification and application of optimized analytical protocols2000 · 185 citations