One-Year Outcomes of the First 1000 Patients Implanted With the Medtronic Micra AV Leadless Pacing System in France: The AV-CESAR Cohort Study

View Full Paper
FKFawzi KerkouriNCN C ClementyPDPascal Defaye

Key Result

The Micra AV leadless pacing system demonstrated a 99.7% successful implantation rate, with early major complications occurring in 2.4% of patients and late pacemaker syndrome in 2.5%.

Key Points

  • This study evaluates the outcomes of the Micra AV leadless pacing system in routine clinical practice.
  • Nationwide retrospective cohort including first 1000 patients implanted with Micra AV from 2020-2024.
  • Mean follow-up duration was 13.4 months with primary endpoints focused on major complications.
  • Data included patient demographics, reasons for implantation, and rates of complications.
  • Device implantation was successful in 1000 of 1003 patients (99.7%).
  • Early major complications occurred in 2.4%; late complications included pacemaker syndrome in 2.5%.
  • All-cause mortality was 13.8%, with only 0.2% being device-related.

Study Design

Type

Cohort (n=1,003)

Multicenter

Yes

Structured PICO

Does the Micra AV leadless pacing system demonstrate acceptable safety and clinical performance in patients requiring pacing in routine clinical practice?

P
Population
1003 patients implanted with a Medtronic Micra AV leadless pacemaker in France. Mean age 72.0±15 years, 62.8% male, 44.0% had ≥2 comorbidities, and 62.4% were unsuitable for transvenous pacemakers. Primary indication was permanent complete atrioventricular block with preserved sinus rhythm (78.0%).
I
Intervention
Medtronic Micra AV leadless pacing system
O
Outcome
Early (in-hospital) and late (postdischarge) device-related major complications; pacemaker syndrome and need for implantation of a new pacing systemsafety

The Micra AV leadless pacemaker demonstrates acceptable safety and performance in a high-risk real-world population, though atrioventricular synchrony declines at higher heart rates.

Abstract

BACKGROUND: Evidence supporting the Micra AV leadless pacing system has largely derived from highly experienced centers, potentially limiting generalizability to routine clinical practice. METHODS: The AV-CESAR study is a nationwide retrospective cohort, including the first 1000 patients implanted with a Medtronic Micra AV leadless pacemaker in France (2020–2024). Mean follow-up was 13.4±10.6 months. Primary end points were early (in-hospital) and late (postdischarge) device-related major complications; pacemaker syndrome and need for implantation of a new pacing system were centrally adjudicated. RESULTS: Among 1003 patients, device implantation was successful in 1000 (99.7%). Mean age was 72.0±15 years; 62.8% were male, 44.0% had ≥2 comorbidities, and 62.4% were unsuitable for transvenous pacemakers. The primary indication was permanent complete atrioventricular block with preserved sinus rhythm (78.0%). Early major complications occurred in 2.4%, including pericardial effusion (0.8%; 3 requiring surgery), access-site complications (0.8%), and thromboembolism (0.3%). Late major complications included pacemaker syndrome (2.5%), pacing-induced cardiomyopathy (1.2%), high thresholds (>4 V/0.24ms, 0.3%), and premature battery depletion (0.2%). No device infections or dislodgements were observed. Overall, 2.5% of patients required implantation of a new pacing system (pacing-induced cardiomyopathy n=11; pacemaker syndrome n=7; high threshold n=3; battery/software failure n=3; tricuspid regurgitation n=1) and 12.7% were permanently reprogrammed to VVI mode. In pacing-dependent patients with ambulatory Holter monitoring (>3.4 million paced cycles), mean atrioventricular synchrony was 67.6±17.3% and correlated with device-reported AM-VP (74.8±17.1%; R 2 =0.92; P 90 beats per minute (5.6% of monitored time). All-cause mortality was 13.8%, with 0.2% directly device-related; 41.7% of deaths were cardiovascular. CONCLUSIONS: The Micra AV system is primarily used as an alternative when dual-chamber conventional pacemakers are not feasible. It demonstrated acceptable safety and clinical performance in a high-risk population. The marked reduction in atrioventricular synchrony at higher heart rates may suggest limited clinical benefit in patients with higher exertional heart rates. REGISTRATION: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT05953558.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Kerkouri et al. (2026) conducted a cohort in Permanent complete atrioventricular block with preserved sinus rhythm (n=1,003). Medtronic Micra AV leadless pacing system was evaluated on Early (in-hospital) and late (postdischarge) device-related major complications. The Micra AV leadless pacing system demonstrated a 99.7% successful implantation rate, with early major complications occurring in 2.4% of patients and late pacemaker syndrome in 2.5%.

synapsesocial.com/papers/6a0ff3d9d674f7c03778cc4dhttps://doi.org/10.1161/circep.126.014850