November 1, 1992Pacing and Clinical Electrophysiology

Electrocardiographically Documented Unnecessary, Spontaneous Shocks in 241 Patients With Implantable Cardioverter Defibrillators

View Full Paper
Ask AI
Bookmark
Share

Population

241 patients who underwent implantable cardioverter defibrillator implantation between March 1983 and…

Design

Cohort

Follow-up

24 ± 20 months

Authors

WGWolfram GrimmBFBelinda F. FloresFMFrancis E. Marchlinski

Discussion

Loading...

Member takes

Overview

AF commonly triggers unnecessary ICD shocks; leaves open whether enhanced detection algorithms reduce events.

Structured PICO

P
Population
241 patients who underwent implantable cardioverter defibrillator (ICD) implantation between March 1983 and November 1991
I
Intervention
Implantable Cardioverter Defibrillator (ICD)
O
Outcome
Incidence and cause of electrocardiographically documented spontaneous ICD discharges for a rhythm other than ventricular tachycardia (VT) or fibrillation (VF) (unnecessary shocks)safety

Unnecessary ICD shocks are a frequent complication occurring in at least 22% of patients, primarily driven by atrial fibrillation, and can be better diagnosed with electrogram storage capabilities.

Cite This Study

Grimm et al. (1992) studied this question.

synapsesocial.com/papers/6a10fb3fcfa01e990d9ff30fhttps://doi.org/10.1111/j.1540-8159.1992.tb02953.x