PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
February 25, 2011Journal of Cardiovascular Electrophysiology14 citations

Is There a Potential Benefit to Increased Irrigation Channels During Radiofrequency Ablation? Results From a Two-Center Prospective Randomized Study

View Full Paper
SKSébastien KnechtFSFrédéric SacherAFAndreï Forclaz

Key Result

The use of a 12-channel irrigated-tip catheter for cavotricuspid isthmus ablation significantly increased the risk of steam pops compared to a 6-channel catheter (13% vs 0%, P=0.018).

Study Design

Type

RCT (n=95)

Multicenter

Yes

Structured PICO

Does a 12-channel irrigated-tip catheter improve procedural efficiency or safety compared to a 6-channel catheter in patients undergoing cavotricuspid isthmus ablation?

P
Population
95 consecutive patients undergoing cavotricuspid isthmus ablation
I
Intervention
Radiofrequency ablation using a 4 mm tip ablation catheter with 12 irrigation channels (6 distal and 6 proximal)
C
Comparator
Radiofrequency ablation using a 3.5 mm tip ablation catheter with 6 distal irrigation channels
O
Outcome
Total procedural duration, radiofrequency duration, fluoroscopic duration, and amount of irrigationsurrogate

The addition of proximal irrigation holes to a radiofrequency ablation catheter does not improve procedural efficiency but significantly increases the risk of steam pops, likely due to distorted temperature feedback.

Main Result

Absolute Event Rate: 13% vs 0%

p-value: p=0.018

Abstract

INTRODUCTION: Open irrigation during radiofrequency (RF) application allows a higher power delivery in the setting of temperature-controlled ablation, without causing blood clots. This study sought to evaluate the clinical value of the additional 6 supplementary channels at the proximal catheter tip compared to a standard irrigated RF catheter with 6 conventional channels present at the distal tip only. METHODS AND RESULTS: Ninety-five consecutive patients were prospectively randomized to cavotricuspid isthmus ablation using an 3.5 mm tip ablation catheter with 6 distal irrigation channels (6C; 48 patients) or an 4 mm tip ablation catheter with 12 irrigation channels (12C; 47 patients) disposed at the distal (6 channels) and proximal (6 additional channels) catheter tip. There was no significant difference between the 12C and the 6C irrigated-tip catheter concerning the total procedural duration, the RF duration, the fluoroscopic duration, and the amount of irrigation. Conversely, there were significantly more patients who experienced at least one steam pop while using the 12C as compared to the 6C irrigated-tip catheter (0% vs 13%, respectively, P = 0.018). CONCLUSION: The addition of proximal irrigation holes at the catheter tip do not facilitate lesion formation during RF ablation, but significantly increases the risk of steam pop. This is probably the consequence of an increase distortion of the temperature feedback.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Knecht et al. (2011) conducted an RCT in Patients undergoing cavotricuspid isthmus ablation (n=95). 12-channel irrigated-tip ablation catheter vs. 6-channel irrigated-tip ablation catheter was evaluated on Incidence of at least one steam pop (p=0.018). The use of a 12-channel irrigated-tip catheter for cavotricuspid isthmus ablation significantly increased the risk of steam pops compared to a 6-channel catheter (13% vs 0%, P=0.018).

synapsesocial.com/papers/6a12fcd1c031bb6829a7a916https://doi.org/10.1111/j.1540-8167.2010.01945.x
Ask AI
Helpful
Bookmark
Share
View Full Paper