Right and left lung perfusion: in vitro and in vivo validation with oblique-angle, velocity-encoded cine MR imaging.

View Full Paper
GCGary R. CaputoCKChisato KondoTMTakayuki Masui

Key Result

Velocity-encoded cine MR imaging enabled accurate and reproducible measurement of absolute blood flow in healthy subjects, correlating well with Doppler echocardiography.

Key Points

  • This research aims to assess the accuracy of velocity-encoded cine MR imaging for measuring pulmonary blood flow.
  • Conducted in vitro validation using a constant-flow phantom.
  • Performed in vivo measurements with nine healthy volunteers.
  • Compared magnetic resonance findings with Doppler echocardiography results.
  • Velocity measurements in the flow phantom correlated well with Doppler echocardiography (p < 0.001).
  • Cine MR provided accurate and reproducible measurements of absolute blood flow in healthy subjects.
  • Oblique-gradient flow encoding was optimal for measuring velocities in the right and left pulmonary arteries.

Study Design

Type

Observational (n=9)

Structured PICO

Does velocity-encoded cine MR imaging accurately measure pulmonary blood flow compared to Doppler echocardiography?

P
Population
In vitro constant-flow phantom and 9 healthy volunteers
I
Intervention
Velocity-encoded cine MR imaging with oblique-angle flow encoding
C
Comparator
Doppler echocardiography and direct measurements
O
Outcome
Accuracy and reproducibility of velocity and volume of blood flow measurements in the main, right, and left pulmonary arteriessurrogate

Velocity-encoded cine MR imaging with oblique-gradient flow encoding is an accurate and reproducible method for measuring absolute blood flow in the pulmonary arteries.

Abstract

Quantification of pulmonary flow is clinically important in the evaluation of both congenital and acquired heart disease. Velocity-encoded cine magnetic resonance (MR) is a promising technique for measuring velocity and volume of blood flow. The authors report validation of the accuracy of velocity-encoded cine MR for measurement of oblique-angle flow in vitro, with use of a constant-flow phantom, and in vivo, with nine healthy volunteers in whom velocities were measured separately in the main, right, and left pulmonary arteries. Findings at MR were compared with findings at Doppler echocardiography. Velocity measurements in a flow phantom with cine MR correlated well with direct measurements at Doppler echocardiography. Velocity-encoded cine MR enabled accurate and reproducible measurement of absolute blood flow in healthy subjects. Oblique-gradient flow encoding (ie, flow-encoding direction coinciding with the true direction of flow) was the method of choice for velocity measurements in the right and left pulmonary arteries.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Caputo et al. (1991) conducted an observational in Healthy volunteers (n=9). Velocity-encoded cine MR imaging vs. Doppler echocardiography was evaluated on Measurement of oblique-angle flow velocity and volume. Velocity-encoded cine MR imaging enabled accurate and reproducible measurement of absolute blood flow in healthy subjects, correlating well with Doppler echocardiography.