Blood Pressure Estimation Using Photoplethysmogram Signal and Its Morphological Features

View Full Paper
NHNavid HasanzadehMAMohammad Mahdi AhmadiHMHoda Mohammadzade

Key Result

A machine learning model using PPG signal features estimated systolic BP with a mean absolute error of 8.22 mmHg (r=0.78) and diastolic BP with a mean absolute error of 4.17 mmHg (r=0.72).

Structured PICO

Does a machine learning model using specific photoplethysmogram (PPG) signal morphological features accurately estimate blood pressure?

I
Intervention
Machine learning model using photoplethysmogram (PPG) signal and morphological features (logarithm of dicrotic notch reflection index, ratio of low- to high-frequency components of heart rate variability, and heart rate multiplied by modified Normalized Pulse Volume)
C
Comparator
Previously reported methods that only use PPG signal
O
Outcome
Accuracy of systolic and diastolic blood pressure estimation (correlation coefficient, mean absolute error, and standard deviation)surrogate

A novel machine learning algorithm using specific PPG morphological features can estimate blood pressure with moderate to high accuracy, achieving BHS Grade A for diastolic BP.

Main Result

Effect estimate: Correlation coefficient 0.78 (SBP), 0.72 (DBP)

Abstract

In this paper, we present a machine learning model to estimate the blood pressure (BP) of a person using only his photoplethysmogram (PPG) signal. We propose algorithms to better detect some critical points of the PPG signal, such as systolic and diastolic peaks, dicrotic notch and inflection point. These algorithms are applicable to different PPG signal morphologies and improve the precision of feature extraction. We show that the logarithm of dicrotic notch reflection index, the ratio of low- to high-frequency components of heart rate (HR) variability signal, and the product of HR multiplied by the modified Normalized Pulse Volume (mNPV) are the key features in accurately estimating the BP using PPG signal. Our proposed method has achieved higher accuracies in estimating BP compared to the previously reported methods that only use PPG signal. For the systolic BP, the achieved correlation coefficient between the estimated values and the real values is 0.78, the mean absolute error of the estimated values is 8.22 mmHg, and their standard deviation is 10.38 mmHg. For the diastolic BP, the achieved correlation coefficient between the estimated and the real values is 0.72, the mean absolute error of the estimated values is 4.17 mmHg, and their standard deviation is 4.22 mmHg. The achieved results fall within Grade A for diastolic, Grade C for systolic and Grade B for mean BP based on BHS standard.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Hasanzadeh et al. (2019) studied Blood pressure estimation. Machine learning model using PPG signal morphological features vs. Real BP values was evaluated on Accuracy of systolic and diastolic BP estimation (mean absolute error and correlation coefficient) (Correlation coefficient 0.78 (SBP), 0.72 (DBP)). A machine learning model using PPG signal features estimated systolic BP with a mean absolute error of 8.22 mmHg (r=0.78) and diastolic BP with a mean absolute error of 4.17 mmHg (r=0.72).