October 26, 2016IEEE Access75 citationsOpen Access

Design of a Real-Time ECG Filter for Portable Mobile Medical Systems

View Full Paper
JLJianqiang LiGDGenqiang DengWWWei Wei

Key Result

An improved integer coefficients IIR digital filter effectively eliminated baseline drift and 50-Hz power frequency interference in ECG signals for portable mobile medical systems.

Structured PICO

P
Population
ECG signals (verified using MIT/BIH Arrhythmia Database and a specific example)
I
Intervention
Improved integer coefficients infinite impulse response (IIR) digital filter
O
Outcome
Elimination of baseline drift and 50-Hz power frequency interference

An improved integer IIR filter effectively removes noise from ECG signals in real-time for portable mobile medical systems.

Abstract

Electrocardiogram (ECG) signal is a direct and effective way to find cardiovascular disease timely, which can intuitively reflect changes of the heart beat and activities of different parts. Due to the noise interference from surroundings, acquisition of real-time and high-quality ECG signal is a big challenge for portable mobile medical systems. Integer coefficients infinite impulse response (IIR) digital filter is suitable for portable mobile platform, but it will be a little bit of distortion. So, an improved integer IIR filter for portable ECG monitors is presented in this paper. The proposed IIR filter can instantaneously and effectively eliminate baseline drift of main interference frequency band and 50-Hz power frequency interference in ECG signal. The method is verified by a specific example and MIT/BIH Arrhythmia Database. The improved integer coefficients IIR filter was applied in a portable mobile medical system. It can meet the requirements of ECG filtering in real-time and filtering performance.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Li et al. (2016) studied Cardiovascular disease (ECG monitoring). Improved integer coefficients IIR digital filter was evaluated on Elimination of baseline drift and 50-Hz power frequency interference. An improved integer coefficients IIR digital filter effectively eliminated baseline drift and 50-Hz power frequency interference in ECG signals for portable mobile medical systems.