Forearm blood flow by Doppler ultrasound during rest and exercise: tests of day-to-day repeatability

View Full Paper
JSJ. Kevin ShoemakerZPZlatko PozegRHRichard L. Hughson

Structured PICO

Are Doppler ultrasound measures of forearm blood flow reproducible across different test days during rest and exercise in volunteers?

P
Population
6 volunteers
I
Intervention
Doppler ultrasound measurement of brachial artery diameter and mean blood velocity during 1-min rest and 4-min dynamic handgrip exercise on 3 separate days
O
Outcome
Between-day repeatability (coefficient of variation) of brachial artery diameter and mean blood velocitysurrogate

Doppler ultrasound provides reproducible day-to-day measurements of brachial artery diameter and mean blood velocity during rest and exercise.

Abstract

The between-day repeatability of simultaneous measures of brachial artery diameter (D) (echo Doppler) and mean blood velocity (MBV) (pulsed Doppler) was tested during rest and exercise. On 3 separate days, six volunteers performed one trial of 1-min rest followed by a step increase in dynamic handgrip exercise for 4 min which required the lifting and lowering of a 4.4-kg weight (approximately 8-12% MVC) in a 1s/2s (work/rest) cadence. Measures for MBV and D were collected continuously on a beat-by-beat basis during the transition from rest to end exercise. The mean rest values over one min, and single data points at 30, 60, 120, and 240 s of exercise were extracted from the time series data. At all exercise time points, MBV was greater than rest (P 0.05), nor did D differ between days. The mean between-day coefficient of variation for D was 4.08 +/- 0.7% at rest and ranged from 2.90 +/- 0.4% to 3.96 +/- 0.5% during exercise. The coefficient of variation for MBV was 13.2 +/- 2.6% at rest and reached 20.2 +/- 3.1% during the final min of exercise; the exercise variability was reduced to 14.9 +/- 2.4% by averaging MBV over 3 s (the duration of a contraction/relaxation duty cycle) (P < 0.05) with no further advantage of averaging over ten 60-s sample periods. The data indicate that, for the six subjects tested, Doppler ultrasound measures of arterial MBV and diameter during both rest and exercise were reproducible across different test days and can be used as a reliable, noninvasive means of testing hypotheses pertaining to blood flow control.