November 1, 1995Hypertension158 citations

State of the Market

View Full Paper
EOEoin OʼBrienNANeil AtkinsJSJan A. Staessen

Structured PICO

Which 24-hour ambulatory blood pressure measurement devices meet the AAMI and BHS criteria for accuracy and performance?

P
Population
43 ambulatory blood pressure measurement devices from 31 manufacturers or suppliers
I
Intervention
Validation according to the protocols of the Association for the Advancement of Medical Instrumentation (AAMI) or the British Hypertension Society (BHS)
O
Outcome
Accuracy and performance according to AAMI or BHS criteriasurrogate

Only a small fraction (9 out of 43) of commercially available 24-hour ambulatory blood pressure monitors have published evidence meeting established accuracy criteria.

Abstract

The introduction of 24-hour ambulatory blood pressure measurement into clinical practice created a large market for ambulatory blood pressure measurement devices. Forty-three such devices from 31 manufacturers or suppliers are now available to satisfy a market demand that is likely to increase. The aim of this article is to identify the devices available and then to examine critically any validation studies assessing accuracy and performance. Of the 43 devices available 18 have been validated according to the protocols of the Association for the Advancement of Medical Instrumentation (AAMI) or the British Hypertension Society (BHS) in 25 reported studies. In 9 of these studies the protocol was not adhered to, and the results, which are therefore questionable, are noted but not considered further. Fourteen devices were evaluated according to the accuracy criteria of both protocols, and of these 9 fulfilled the requirements. From this review of 43 devices on the market it may be concluded that, at the time of writing, there is published evidence for only 9 devices meeting the generally accepted AAMI and BHS criteria for accuracy and performance; these are the A&D TM-2420 models 6 and 7 and TM-2421, CH-Druck, Nissei ABPM DS-240, Profilomat, QuietTrak, and SpaceLabs SL-90202 and SL-90207.