Noninvasive Measurement of Medium-Sized Artery Intima-Media Thickness in Humans: In vitro Validation

View Full Paper
XGXavier GirerdJMJean‐Jacques MouradCAChristophe Acar

Structured PICO

Does a high-resolution echo-tracking device accurately measure radial artery intima-media thickness compared to histology in patients with coronary heart disease?

P
Population
15 radial artery segments obtained from the distal end of the wrist-elbow harvest for coronary bypass grafting in patients with coronary heart disease. Subgroups included 11 patients for in vivo vs in vitro comparison and 10 subjects for repeatability.
I
Intervention
High-resolution echo-tracking device (ultrasound) for measuring radial artery intima-media thickness
C
Comparator
Histological measurements
O
Outcome
Correlation between histological and ultrasonic measurements of intima-media thicknesssurrogate

A high-resolution echo-tracking device provides reasonably accurate measurements of radial artery intima-media thickness compared to histology, supporting its use for noninvasive arterial assessment.

Abstract

Recent research in ultrasound technology has led to the development of a high-resolution echo-tracking device. The present study was performed to evaluate the accuracy in the measurement of human radial artery intima-media thickness with this new device. We determined the correlation between histological and ultrasonic measurements of intima-media thickness in 15 radial artery segments obtained from the distal end of the wrist-elbow harvest for coronary bypass grafting in patients with coronary heart disease. For arterial intima-media thickness, a positive correlation was observed between ultrasonic and histological measurements (r = 0.618; p < 0.014), and the difference between ultrasound and histology measurements was 41 +/- 66 microns, with the higher measurements found by the ultrasonic device. In a subgroup of 11 patients, we determined the correlation between in vivo ultrasonic measurements of radial artery intima-media thickness at the preoperative stage and in vitro ultrasonic measurements of intima-media thickness obtained postoperatively in the same arterial segments. Internal diameter was larger in vivo than in vitro, and intima-media thickness was smaller in vivo than in vitro. The cross-sectional area of the arterial wall was calculated from internal diameter and intima-media thickness. In vitro wall cross-sectional area was correlated with in vivo wall cross-sectional area (r = 0.929; p < 0.0001). Repeatability of in vivo intima-media thickness measurements was investigated in 10 subjects through the calculation of the repeatability coefficient as defined by the British Standards Institution.(ABSTRACT TRUNCATED AT 250 WORDS)