Sequential Biventricular Pacing:

View Full Paper
PMPeter MortensenPSPeter SøgaardHMHassan Mansour

Key Result

Sequential biventricular pacing significantly improved stroke volume compared to simultaneous stimulation (68 vs 56 mL, P<0.001) and improved 6-minute walk distance and NYHA class at 3 months.

Study Design

Type

Observational (n=198)

Multicenter

Yes

Structured PICO

Does sequential biventricular pacing improve stroke volume and clinical symptoms compared to simultaneous pacing in patients with symptomatic systolic heart failure and prolonged QRS?

P
Population
198 patients with symptomatic systolic heart failure and a prolonged QRS complex duration
I
Intervention
Sequential biventricular pacing using the InSync III (Model 8042) biventricular stimulator
C
Comparator
Simultaneous biventricular pacing (for stroke volume comparison) and baseline status (for clinical metrics)
O
Outcome
Clinical safety, performance, and efficacy (including 6-minute walk distance, NYHA class, and stroke volume) at 3 monthssurrogate

Sequential biventricular pacing is safe and provides superior stroke volume compared to simultaneous biventricular pacing in patients with systolic heart failure and prolonged QRS.

Main Result

Absolute Event Rate: 68% vs 56%

p-value: p=<0.001

Abstract

The study evaluated the clinical safety, performance, and efficacy of sequential biventricular pacing in the InSync III (Model 8042) biventricular stimulator in a multicenter, prospective 3-month study and assessed the proper functioning of features aiming at improving biventricular AV therapy delivery. The system was successfully implanted in 189 (95.9%) of 198 patients with symptomatic systolic heart failure and a prolonged QRS complex duration. Patients significantly improved their 6-minute hall walk distance (baseline 339 +/- 92 vs 3-month 422 +/- 127 meter, P < 0.001) and NYHA class (baseline 3.1 +/- 0.5 vs 3-month 1.9 +/- 0.7, P < 0.001). Echocardiographic optimization of sequential biventricular pacing showed an improvement in stroke volume compared to simultaneous stimulation (sequential 68 +/- 24 mL vs simultaneous 56 +/- 23 mL, P < 0.001) at baseline and at 3 months. In 88% (30/34) of the patients these improvements were seen within a small range of V-V delays of +/-20 ms and in 94% (32/34) within V-V delays of +/-40 ms. In contrast, programming beyond this range reduced stroke volume below that during simultaneous biventricular pacing. The device functioned as expected. LV lead dislodgement was observed in 12 patients and phrenic nerve stimulation required lead repositioning in 2 patients. Eight patients died during the study. Patient survival at 3 and 6 months was 97 +/- 2% and 94 +/- 2%, respectively. Cause of death was cardiac (n = 7), heart failure related (n = 3), arrhythmia related (n = 2), and unknown (n = 2). In conclusion, this sequential biventricular pacemaker was safe and efficacious.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Mortensen et al. (2004) conducted an observational in Symptomatic systolic heart failure and a prolonged QRS complex duration (n=198). Sequential biventricular pacing vs. Simultaneous biventricular pacing was evaluated on Stroke volume (p=<0.001). Sequential biventricular pacing significantly improved stroke volume compared to simultaneous stimulation (68 vs 56 mL, P<0.001) and improved 6-minute walk distance and NYHA class at 3 months.