February 12, 2016Journal of Magnetic Resonance Imaging46 citationsOpen Access

Assessment of the precision and reproducibility of ventricular volume, function, and mass measurements with ferumoxytol‐enhanced 4D flow MRI

KHKate HannemanAKAya KinoJCJoseph Y. Cheng

Key Result

Ferumoxytol-enhanced 4D flow MRI quantified ventricular volume, function, and mass with precision and interobserver agreement comparable to cine SSFP (r=0.808-0.972 for volumes, P<0.001).

Study Design

Type

Observational (n=22)

Blinding

Blinded

Structured PICO

Does 3D time-resolved (4D) flow MRI provide comparable precision and interobserver agreement to cine SSFP for quantifying ventricular volume, function, and mass in patients with congenital heart disease?

P
Population
22 consecutive patients with congenital heart disease (CHD) (10 males, 6.4 ± 4.8 years) referred for 3T ferumoxytol-enhanced cardiac MRI
I
Intervention
3D time-resolved (4D) flow MRI (ferumoxytol-enhanced)
C
Comparator
2D short-axis cine steady-state free precession (SSFP)
O
Outcome
Precision and interobserver agreement of ventricular volume, function, and mass quantificationsurrogate

4D flow MRI provides comparable precision and interobserver agreement to standard cine SSFP for quantifying ventricular volume, function, and mass in patients with congenital heart disease.

Main Result

Effect estimate: r = 0.808-0.972

p-value: p=< 0.001

Abstract

PURPOSE: To compare the precision and interobserver agreement of ventricular volume, function, and mass quantification by 3D time-resolved (4D) flow MRI relative to cine steady-state free precession (SSFP). MATERIALS AND METHODS: With Institutional Research Board approval, informed consent, and HIPAA compliance, 22 consecutive patients with congenital heart disease (CHD) (10 males, 6.4 ± 4.8 years) referred for 3T ferumoxytol-enhanced cardiac MRI were prospectively recruited. Complete ventricular coverage with standard 2D short-axis cine SSFP and whole chest coverage with axial 4D flow were obtained. Two blinded radiologists independently segmented images for left ventricular (LV) and right ventricular (RV) myocardium at end systole (ES) and end diastole (ED). Statistical analysis included linear regression, analysis of variance (ANOVA), Bland-Altman (BA) analysis, and intraclass correlation (ICC). RESULTS: Significant positive correlations were found between 4D flow and SSFP for ventricular volumes (r = 0.808-0.972, P < 0.001), ejection fraction (EF) (r = 0.900-928, P < 0.001), and mass (r = 0.884-0.934, P < 0.001). BA relative limits of agreement for both ventricles were between -52% to 34% for volumes, -29% to 27% for EF, and -41% to 48% for mass, with wider limits of agreement for the RV compared to the LV. There was no significant difference between techniques with respect to mean square difference of ED-ES mass for either LV (F = 2.05, P = 0.159) or RV (F = 0.625, P = 0.434). Interobserver agreement was moderate to good with both 4D flow (ICC 0.523-0.993) and SSFP (ICC 0.619-0.982), with overlapping confidence intervals. CONCLUSION: Quantification of ventricular volume, function, and mass can be accomplished with 4D flow MRI with precision and interobserver agreement comparable to that of cine SSFP. J. Magn. Reson. Imaging 2016;44:383-392.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Hanneman et al. (2016) conducted an observational in congenital heart disease (CHD) (n=22). 3D time-resolved (4D) flow MRI vs. cine steady-state free precession (SSFP) was evaluated on ventricular volume, function, and mass quantification (r = 0.808-0.972, p=< 0.001). Ferumoxytol-enhanced 4D flow MRI quantified ventricular volume, function, and mass with precision and interobserver agreement comparable to cine SSFP (r=0.808-0.972 for volumes, P<0.001).