April 3, 2013IEEE Journal of Solid-State Circuits187 citations

A Low-Power Processor With Configurable Embedded Machine-Learning Accelerators for High-Order and Adaptive Analysis of Medical-Sensor Signals

View Full Paper
KLKyong-Ho LeeYonsei UniversityNVNaveen VermaPrinceton University

Key Result

A custom processor with embedded machine-learning accelerators reduced energy consumption for EEG-based seizure and ECG-based cardiac-arrhythmia detection by 62.4× and 144.7× compared to a CPU.

Structured PICO

Does a custom processor with embedded machine-learning accelerators reduce energy consumption for medical sensor signal analysis compared to a CPU-based implementation?

P
Population
Clinical data for EEG-based seizure detection and ECG-based cardiac-arrhythmia detection
I
Intervention
Custom processor integrating a CPU with configurable machine-learning accelerators (support-vector-machine and embedded active learning)
C
Comparator
Implementation based on the CPU
O
Outcome
Energy consumption per detectionsurrogate

A custom processor with embedded machine-learning accelerators significantly reduces energy consumption for medical sensor signal analysis, including ECG-based arrhythmia detection.

Abstract

Low-power sensing technologies have emerged for acquiring physiologically indicative patient signals. However, to enable devices with high clinical value, a critical requirement is the ability to analyze the signals to extract specific medical information. Yet given the complexities of the underlying processes, signal analysis poses numerous challenges. Data-driven methods based on machine learning offer distinct solutions, but unfortunately the computations are not well supported by traditional DSP. This paper presents a custom processor that integrates a CPU with configurable accelerators for discriminative machine-learning functions. A support-vector-machine accelerator realizes various classification algorithms as well as various kernel functions and kernel formulations, enabling range of points within an accuracy-versus-energy and -memory trade space. An accelerator for embedded active learning enables prospective adaptation of the signal models by utilizing sensed data for patient-specific customization, while minimizing the effort from human experts. The prototype is implemented in 130-nm CMOS and operates from 1.2 V-0.55 V (0.7 V for SRAMs). Medical applications for EEG-based seizure detection and ECG-based cardiac-arrhythmia detection are demonstrated using clinical data, while consuming 273 μJ and 124 μJ per detection, respectively; this represents 62.4× and 144.7× energy reduction compared to an implementation based on the CPU. A patient-adaptive cardiac-arrhythmia detector is also demonstrated, reducing the analysis-effort required for model customization by 20 ×.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Lee et al. (2013) studied Seizure and cardiac-arrhythmia detection. Custom processor with configurable embedded machine-learning accelerators vs. CPU-based implementation was evaluated on Energy consumption per detection. A custom processor with embedded machine-learning accelerators reduced energy consumption for EEG-based seizure and ECG-based cardiac-arrhythmia detection by 62.4× and 144.7× compared to a CPU.

synapsesocial.com/papers/6a1c13bd27b545b111a96dcahttps://doi.org/10.1109/jssc.2013.2253226