Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium

View Full Paper
HPHelen Høgh PetersenXCXu ChenAPAdrian Pietersen

Structured PICO

How do catheter-tip length, ablation site, and convective cooling affect lesion dimensions during temperature-controlled radiofrequency ablation in porcine myocardium?

P
Population
In vitro strips of porcine left ventricular myocardium and in vivo pig hearts
I
Intervention
Temperature-controlled radiofrequency ablation with 2- to 12-mm-tip catheters
C
Comparator
Different catheter-tip lengths, ablation sites (apex vs free wall/septum), and levels of convective cooling
O
Outcome
Lesion dimensions (volume)surrogate

Longer electrode tips increase lesion size up to 6-8 mm, and tip temperature achieved is a poor predictor of lesion size during temperature-controlled ablation.

Abstract

BACKGROUND: It is important to increase lesion size to improve the success rate for radiofrequency ablation of ischemic ventricular tachycardia. This study of radiofrequency ablation, with adjustment of power to approach a preset target temperature, ie, temperature-controlled ablation, explores the effect of catheter-tip length, ablation site, and convective cooling on lesion dimensions. METHODS AND RESULTS: In vitro strips of porcine left ventricular myocardium during different levels of convective cooling and in vivo pig hearts at 2 or 3 left ventricular sites were ablated with 2- to 12-mm-tip catheters. We found increased lesion volume for increased catheter-tip length </=8 mm in vitro (P<0.05) and 6 mm in vivo (P<0. 0001), but no further increase was found for longer tips. For the 4- to 10-mm catheter tips, we found smaller lesion volume in low-flow areas (apex) than in high-flow areas (free wall and septum) (P<0.05). Increasing convective cooling of the catheter tip in vitro increased lesion volume (P<0.0005) for the 4- and 8-mm tips but not for the 12-mm tip as the generator reached maximum output. In contrast to power-controlled ablation, we found a negative correlation between tip temperature reached and lesion volume for applications in which maximum generator output was not achieved (P<0. 0001), whereas delivered power and lesion volume correlated positively (P<0.0001). CONCLUSIONS: Lesion size differs in different left ventricular target sites, which is probably related to convective cooling, as illustrated in vitro. Longer electrode tips increase lesion size for tip lengths </=6 to 8 mm. For temperature-controlled ablation, the tip temperature achieved is a poor predictor of lesion size.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Petersen et al. (1999) studied this question.

synapsesocial.com/papers/6a1c4aec1567d2fc4d5fe445https://doi.org/10.1161/01.cir.99.2.319
Ask AI
Helpful
Bookmark
Share
View Full Paper